Alarm/Indicator:
Device status:
Indicator light/
audio:
Impact to therapy:
Steps to remedy:
High Flow / Leak Alarm
Device detects a significant
air leak.
Status/Alarm
Indicator:
Solid Yellow
The audible alarm
will sound.
The audible alarm
can be paused
by pressing the
Audio Pause
button. When
pressed the light
above the button
will illuminate.
Device will continue
to operate, but may
not provide the
prescribed therapy.
• Press down around the
edges of your dressing. If
your dressing begins to
recompress, you have
resealed your dressing and
addressed the air leak.
• Check the connectors on
the tubing between your
dressing and the canister.
Ensure the connection is
secure (see images on pages
20 and 22.)
• Disconnect the quick click
connectors and close the
tethered caps to seal the
tubing. If the alarm continues,
replace the canister.
• If you cannot locate the
source of your air leak, contact
your healthcare provider.
Содержание RENASYS GO 66801496
Страница 2: ...Healthcare provider s name Healthcare provider s telephone number Healthcare provider s mobile number...
Страница 86: ...86 FR Page intentionaly left blank...
Страница 128: ...128 DE Page intentionaly left blank...
Страница 170: ...170 ES Page intentionaly left blank...
Страница 212: ...212 PT Page intentionaly left blank...
Страница 254: ...254 IT Page intentionaly left blank...
Страница 296: ...296 NL Page intentionaly left blank...
Страница 338: ...338 DA Page intentionaly left blank...
Страница 380: ...380 NO Page intentionaly left blank...
Страница 422: ...422 FI Page intentionaly left blank...
Страница 464: ...464 SV Page intentionaly left blank...
Страница 477: ...477 TR sm ith nep hew REN ASY S G O Negat i v e Pressu re Woun d Therap y G jak G G kablosu giri i DC kablosu G kayna...
Страница 512: ......