29
EN
Alarm/Indicator:
Device status:
Indicator light/
audio:
Impact to therapy:
Steps to remedy:
Device is not plugged into an
electrical outlet and battery
is extremely low.
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
will sound.
The audible
alarm cannot be
paused.
The battery has
only
2 minutes
of therapy times
remaining. Device
must be plugged
into electrical outlet
to continue therapy.
If it is not plugged
into an electrical
outlet it will stop
operating once
the battery is fully
discharged.
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.
Device is not plugged into
an electrical outlet and
device battery is extremely
low. After 2 minutes in an
extremely low charge state,
the device will power off.
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
The audible alarm
will sound.
The audible
alarm cannot be
paused.
Device must be
plugged into an
electrical outlet to
continue therapy.
Plug the device into an
electrical outlet.
Therapy will continue to be
delivered while the device
battery is charging.
Device is plugged into an
electrical outlet and battery
is not charging.
Note:
Device display and
indicator lights will only
present the alarm state
when the device is plugged
into an electrical outlet.
Status/Indicator:
Solid Yellow
Battery Status
Bottom Indicator:
Solid Yellow
Device must be
plugged into an
electrical outlet to
continue therapy.
If it is not plugged
into an electrical
outlet it will stop
operating once
the battery is fully
discharged.
Contact your healthcare
provider if the use of a battery
is preferred.
Содержание RENASYS GO 66801496
Страница 2: ...Healthcare provider s name Healthcare provider s telephone number Healthcare provider s mobile number...
Страница 86: ...86 FR Page intentionaly left blank...
Страница 128: ...128 DE Page intentionaly left blank...
Страница 170: ...170 ES Page intentionaly left blank...
Страница 212: ...212 PT Page intentionaly left blank...
Страница 254: ...254 IT Page intentionaly left blank...
Страница 296: ...296 NL Page intentionaly left blank...
Страница 338: ...338 DA Page intentionaly left blank...
Страница 380: ...380 NO Page intentionaly left blank...
Страница 422: ...422 FI Page intentionaly left blank...
Страница 464: ...464 SV Page intentionaly left blank...
Страница 477: ...477 TR sm ith nep hew REN ASY S G O Negat i v e Pressu re Woun d Therap y G jak G G kablosu giri i DC kablosu G kayna...
Страница 512: ......