22
Specifications
Pressure
Setting
Maximum
40, 50, 60, 70, 80, 90, 100, 120, 140, 160, 180 and 200mmHg:
tolerance ± 10mmHg
200mmHg
Alarms
General (all alarms)
Priority
Auditory sound level
Low
46dB
Over vacuum
Indicator color
Yellow
High flow/leak
Indicator color
Alarm overall delay
Yellow
60 seconds ± 5 seconds
Low vacuum
Indicator color
Alarm overall delay
Yellow
30 seconds ± 5 seconds
High vacuum
Indicator color
Alarm overall delay
Yellow
30 seconds ± 5 seconds
Blockage/canister full
Indicator color
Alarm overall delay
Yellow
120 seconds ± 5 seconds
Low battery
Indicator color
Yellow
Battery failed
Indicator color
Yellow
Inactive
Indicator color
Yellow
Device not calibrated
Indicator color
Yellow
Device failed
Indicator color
Yellow
Содержание RENASYS GO 66801244