25
M71 Pressure
Measure
Pressure acquisition mode setting.
Options: manual, or auto.
Manual: If select manual, the
pressure value will be replaced by the
manual setting value (the value set in
basic menu “Gauge Pre. Kpa”).
Auto: If select auto, pressure value is
by automatic acquisition, need to
connect with external pressure
sensor.
Menu
Selection
M72
Temperatur
e
Low Trim
Enter the lower calibration resistance
value.
unit: ohm.
Use standard resistance as input.
For example: 1000 for Pt1000, or 100
for Pt100.
Direct
Input
M73
Temperatur
e
Low Trim
Enter the high calibration resistance
value.
unit: ohm.
Use standard resistance as input.
For example: 2500 for Pt1000, or 250
for Pt100.
Direct
Input
M74 Pressure
Low Trim
Enter the calibration reference
pressure value.
Unit: KPa.
Apply the standard pressure to the
sensor.
For example: 0 KPa
Direct
Input
M75 Pressure
High Trim
Enter the calibration reference
pressure value.
Unit: KPa.
Apply the standard pressure to the
sensor.
For example: 1000 KPa
Direct
Input