24
M62
AMP.
Channel
There are CH_1, CH_2, CH_3 three
options.
CH_3 gain maximum
CH_1 gain minimum
Note:
CH2 generally used for liquid
measurement, which corresponds to
the configuration software, select X1
and X2.CH_3 generally used for gas
measurement, which corresponds to
the configuration software, select X1,
X2 and X3.
****62 Menu
Selection
M63
Work
Mode
There are F_1, F_2, F_3, F_4 four
options.
F_1: Anti-vibration Mode
F_2: Normal Mode
F_3: Turbine Mode
F_4: Test Mode
Note:
Generally choose F_2.
****63 Menu
Selection
M40 Trim 4mA
Steps:
1.Press M-Key for 2 seconds, enter
trim;
2.Short press M-key to decrease
current. Press S-Key to increase
current. Stepping is 16 microamperes.
3.Press M-Key for 2 seconds to save
new trim value.
Or press Z-Key to exit without saving.
****40
M41 Trim 20mA
M70 Temp.
Measure
Temperature acquisition mode
setting.
Options: Manual, or Auto.
Manual: Temperature uses the input
reference value.
Auto: Temperature is automatic
acquisition, should be use external
Pt1000 or Pt100.
****70
Menu
Selection