HRX-OM-Y120
Chapter 3 Serial communication
3.13
Register Map
HRL Series
3-26
●
Function code 4 : Read one data of 0038 h
(Read out the ambient temperature of data display 1.)
Request message 010400380001
Field name
Example
(HEX)
ASCII mode
character data
RTU mode
binary data
Header
-
“
:
”
None
Slave Address
0x01
“0”,“1”
0x01
Function
0x04
“0”,“4”
0x04
Head address of specified register
0x0038
“0”,“0”,“3”,“8”
0x00,0x38
Quantity of register to read
0x0001
“0”,“0”,“0”,“1”
0x00,0x01
Checksum ASCII:LRC RTU:CRC
-
“C”,”2”
0xB0,0x07
Trailer
-
CR/LF
None
Total
quantity of byte
17
8
Response 01040200FA
Field name
Example
(HEX)
ASCII mode
character data
RTU mode
binary data
Header
-
“
:
”
None
Slave Address
0x01
“0”,“1”
0x01
Function
0x04
“0”,“4”
0x04
Quantity of bytes to read
0x02
“0”,“2”
0x02
Information of 0038h (Data display
1)
0x00FA
“0”,“0”,“F”,“A”
0x00,0xFA
Checksum ASCII:LRC RTU:CRC
-
“F”,”F”
0x39,0x73
Trailer
-
CR/LF
None
Total
quantity of byte
15
7
Response message at an ambient temperature of 25.0
℃