Model 5100-XX-IT Toxic Gas Sensor Module
Page: 29
7. APPENDICES
APPENDIX A - SPECIFICATIONS
Sensor type:
Electrochemical
5100-03-IT O2
5100-04-IT CO
5100-05-IT H2S
5100-06-IT Cl2
Max. Range
0-25 %Vol
0-1000 PPM
0-100 PPM
0-10 PPM
User Settings Range
0-100, 0-500, variable
0-10, variable
0-5, variable
Zero Drift
+/- 0.1%
+/- 1 PPM
+/- 0.5 PPM
+/- 0.5 PPM
Repeatability
+/- 0.2%
+/- 1 PPM
+/- 1 PPM
+/- 0.5 PPM
Linearity
+/- 0.2%
+/- 1 PPM
+/- 1 PPM
+/- 0.5 PPM
Resolution
+/- 0.1%
0.5 PPM
0.1 PPM
0.1 PPM
Response Time (1)
<10 sec
<25 sec.
<30 sec.
<60 sec
Accuracy
+/- 0.2%
+/- 1 PPM
+/- 0.5 PPM
+/- 0.5 PPM
Sensor Life
2 years
2 years
2 years
2 years
(1) Response time to 90% full signal value for applied concentration
(2) Sensor life typical for use at standard temperature and pressure with occasional exposure to gas of interest
Output:
Display:
Fixed and Scrolling LED
Relays (Optional):
8 Amp, High Alarm, Low Alarm, 2 Amp Trouble all SPDT
Analog Output (Optional)
Analog 4-20 mA (Trouble 0 mA, Calibration 1.5 mA)
Signal Output:
Sentry bus
Serial RS-485 Modbus RTU
Input:
Remote Alarm Reset:
Normally open digital input (supervised or non-supervised)
Electrical Data:
Power consumption:
2 watts
Connection type:
3 wire
Input voltage:
24 VDC nominal: 10-30VDC
Operating Range:
5100-03-IT O2
5100-04-IT CO
5100-05-IT H2S
5100-06-IT Cl2
Ambient Temp (
o
F)
5 to 122
o
F
-4 to 122
o
F
-40 to 122
o
F
-4 to 122
o
F
Ambient Temp (
o
C)
-15 to 50
o
C
-20 to 50
o
C
-40 to 50
o
C
-20 to 50
o
C
Relative Humidity
5 – 99%
5 – 99%
5 – 99%
5 – 99%
Enclosure:
Material - Standard:
Polyester powder-coated, sand-cast, copper-free aluminum
Material - Stainless Steel Option
316 Stainless Steel (FM approval pending)
Dimensions:
7.5 x 5.0 x 4.8 inches Al, 7.5 x 5.0 x 4.4 inches SS
(H x W x D)
(21.6 x 12.7 x 12.2 cm)
Weight:
2.7 lb. (1.3 Kg) (standard Aluminum enclosure)
5.4 lb. (2.6 Kg) 316 Stainless Steel
Housing:
NEMA
4X
Electrical Classification
Performance:
Explosion proof, Class 1, Div. I, Groups B, C, D
Class I, Zone 1, groups IIB+H2 IP66
Approvals:
FM Approved
Performance and Safety Approval Pending
Содержание 5100 IT Series
Страница 2: ...Contents THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 10: ...Page 6 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 16: ...Page 12 Figure 3 4 4 20mA Circuit Types...
Страница 17: ...Page 13 Figure 3 5 4 20mA Circuit Type Connections for 5100 XX IT...
Страница 18: ...Page 14 For end of line refer to Figure 3 3 Figure 3 6 Wiring Connections for Modbus and Sentry Interface...
Страница 20: ...Page 16 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 27: ...Model 5100 XX IT Toxic Gas Sensor Module Page 23 THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 37: ...Model 5100 XX IT Toxic Gas Sensor Module Page 33 APPENDIX D REMOTE SENSOR DISPLAY DRAWING 5394 52...