ELECTROMAGNETIC COMPATIBILITY
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device
should assure that it is used in such an environment.
±8 kV contact
±15 kV air
power supply lines: ±2 kV
input/output lines: ±1 kV
line(s) to line(s): ±1 kV
line(s) to earth: ±2 kV 100
kHz repetition frequency
0%UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°,225°,270° and
315°
0%UT ; 1 cycle
and
70%UT ; 25/30 cycles
Single phase: at 0°
0% UT ; 300 cycle
30 A/m
50Hz/60Hz
±8 kV contact
±15kV air
power supply lines:
±2 kV
line(s) to line(s): ±1 kV
100 kHz repetition
frequency
0% UT ; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°,225°,270° and
315°
0% UT ; 1 cycle
and
70% UT ; 25/30 cycles
Single phase: at 0°
0% UT ;300 cycle
30 A/m
50Hz/60Hz
Electrostatic Discharge
IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interuptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50Hz/60Hz)
magnetic field
IEC 61000-4-8
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic environment
guidance
Floors should be wood, oncrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be
at least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.