TROUBLE SHOOTING
Problem
Possible Cause
Solution
Cuff position was not correct
or was not tight enough.
Body posture was not correct
during measurement.
Speaking, excessive
movement, angry, excited
or nervous anxiety during
measurement.
Batteries may be expired or
installed improperly.
Incorrect operation; possible
strong electromagnetic
interference.
Display shows an
abnormal result.
Display is blank
when power is on.
No response when
STOP/START is
pressed or when
new batteries have
been loaded.
Apply the cuff properly and attempt
a new measurement.
Review the ‘Taking a Measurement’
portion of this manual and attempt
a new measurement.
Wait a period of time and attempt
a new measurement after calm has
returned; do not speak or move
during measurement.
Check the polarity of the batteries
and reinstall if necessary; replace
batteries with new batteries.
Remove batteries and wait a
minimum of five minutes; reinstall
batteries and attempt a new
measurement.