Smart Sensor Model AS850 Скачать руководство пользователя страница 6

-03-

-12-

Keypad

1.4 Specification
Display: 4-digital LCD display
Minimum display unit: 0.1 mm
Working frequency: 5MHz
Measuring range: 1.2 to 225.0mm (steel)
Minimum limit for tube measuring: 

Φ

20*3mm (steel)

Accuracy: +/- (1%H+0.1)mm,H denotes the measured thickness. 
Sound velocity  range: 1000 to 9999 m/s
Measuring sound velocity with a given thickness: measuring range:
1000 to 9999 m/s. When the given thickness over 20mm, the accuracy
is +/-5%; when the given thickness less than 20mm, the 
accuracy is +/-1mm/H*100%.
Operation temperature: 0

 to 40

 

Power supply: 3*1.5V AAA alkaline batteries
Operation current: Normal operation current

50mA

                                 With Backlight turn on current

120mA

                                 Stand-by current: 

20uA

Size: 70.0*145.5*28mm

1.5 Features
       Auto calibration to assure the accuracy.
       Auto linear compensation: this advanced software program enha
nces the precision by correcting the non-linear accuracy of transducer. 

Take care for measurement while measuring the known rusting spot/
suspicious area. or using sound insulation boardcelotex to locates the
spot in different testing angles.

5.3 Identify different velocity with vary material
    A fault reading would obtains, when measuring the hardware with the
velocity calibrated by prior material. So a correct velocity should be ad-
opted. The fault reading may also result form the difference between
the actual velocity with the calibrated value. 

5.4 Abrasion fo the transducer
Because the transducer is made of propylene, long period use will cause
the surface of transducer became more rough which will decline the sen-
sitivity lead to the wrong reading. Please polish the surface with sand
paper or whetstone to assure the smoothness and parallel. If the reading
still unsteady, the transducer should be replaced with new one.

5.5  “ CAL ” function
    “ CAL ”(calibration) is used to calibrate the unit with the standard block
on the panel, do press this key for calibration with other materials or the 
wrong measuring will take place.

5.6 Multilayer / composite material
It is impossible to read out the thickness of the uncoupled multilayer for
the ultrasonic wave can not go through the uncoupled space. Further
more, the sonic wave cannot travel in the composite material at an even
speed, so ultrasonic reflect principle cannot be applied for measuring
the multilayer/composite material.

5.7 Influence from the oxidized surface
    For some metals, such as aluminum a layer of oxide being generated 
on their surface. The oxidized layer combined with the substrate tightly,
but the sonic wave travel within 2 different material which will lead to 
error reading, the more oxidized layer the reading will be more tolerant.
Please calibrated the unit with the sample block that pick up along the
hardware to be measured, and obtain the thick of sample block by using

4. Keypad diagram
  ON/OFF -- ON/OFF key
  CAL  -- Calibration key
  VEL -- Sound velocity key
  STORE  --  Measurement 
                      data stored key
  CAL+ON/OFF --Backlight active keys
          --Sound velocity, thickness,
thickness unit adjust/recall key
         --Sound velocity, thickness,
thickness unit adjust/recall key

VAL

LIGHT

CAL

STORE

Содержание Model AS850

Страница 1: ...Ultrasonic Thickness Gauge Instruction Manual Model AS850 ...

Страница 2: ......

Страница 3: ...ty due to transportation damages incorrect use or operation manipulation alterations or repair attempts without warranty card invioce 7 2 Non warranty list LCD battery probe sample block plastic case coupling agent 8 Table of sound velocity Sound velocities of common materials Aluminum Zinc Silver Glod Tin Iron Steel Brass Copper SUS Acrylic resin Water 20 Glycerinl soluble glass 2670 3530 4430 54...

Страница 4: ...arget material See the diagram below 01 14 Hardware Display driver Transmitting Receiving Feedback wave C P U LCD display Keypad Counter Hi frequency oscillator 6 3 Cleaning the cabinet Do not use solvent alcohol for cleaning which erode the cabinet LCD window brush and sweep only with a moist cotton cloth 6 4 Cleaning the sample block Because of coupling agent should be put on the sample block du...

Страница 5: ...he coupling agent should be used in proper way typic cally a single droplet of agent is sufficient It is important to use proper coupling agent low viscosity agent the provided agent machining oil is suitable for smooth surface For rough veritcal aluminum surface high viscosity agent like glycerin and lubrication grease is applicable All kinds of coupling agent is available in local market you can...

Страница 6: ...e of propylene long period use will cause the surface of transducer became more rough which will decline the sen sitivity lead to the wrong reading Please polish the surface with sand paper or whetstone to assure the smoothness and parallel If the reading still unsteady the transducer should be replaced with new one 5 5 CAL function CAL calibration is used to calibrate the unit with the standard b...

Страница 7: ... the sound velocity slight different To reach the exact reading the referring block has similar structure as the hardware In comment measurement you check the sound velocity on the table stated on this manual instead of taking calibration for the referring block However this table just for reference sometimes the sound vel ocity will be different cause by different physical chemical factors The so...

Страница 8: ...axis of pipe perpendicular another is parallel and the smaller of two displayed values should then be taken as the thickness at that point 4 5 Complex shape material For complex shape material measurement please refer to the 4 4 the smaller of the two reading should then be taken as the thickness 4 6 Non parallel surface To get a satisfying ultrasonic response the surface must have its one measuri...

Страница 9: ...OFF to turn off the unit If no any operation for 2 minutes the gauge will turn off automaticly and the back light function will be cancelled 4 Measurement Tips 4 1 Cleaning surface Before measuring the dust dirt rusting and grease etc that adheres on the hardware workpiece must be removed off and cleaned 4 2 Decreasing the roughness of surface Too rough surface may result in measure error fault re...

Страница 10: ...elocity and save in current velocity memory unit Measuring the thickness Adjusting actual thickness Velocity read out 1 Keep pressing STORE for 2 seconds to enter save thickness data mode LCD display THICKNESS mm icon with first memory unit If the first memory unit is not being registered so LCD will display 0 0 2 Press or to select your desired memory unit 1 10 3 After picked up the memory unit i...

Отзывы: