Positioning the model on the object holder
Positioning:
The correct positioning of the model on the object holder is indispensible for correct measuring
results. The positioning must be maintained in respect to the "height alignment".
Height alignment:
The scanner has a definite height measuring range. If the object to be measured is located outside
this range, no measuring values will be measured there. This means that the finished scan will
exhibit holes or have a cut-off effect.
The scanner has a height measuring range of +/-25 mm, i.e. 50 mm in total.
Incorrect height alignment is one of the most frequent reasons for unsatisfactory
scanning results!
Figure 36
: Height alignment of Activity850