Figure 21Individual calibration block
These values must be saved in the software after reinstalling the software or replacing the
calibration model.
Start this procedure as follows:
A window opens under Options
-7
Service
-7
Calibration object registration (Fig. 22).
Figure 22: Calibration model registration
Individual values are indicated on the rear of the calibration model. Please enter these in fields #1
and #2.
Confirm the procedure with "OK".
You can start the 3D calibration only by selecting the start command under the item Options
-7
Service
-7
3D calibration
-7
Complete.
The system now performs a 3D calibration; this can take several minutes.
The following message will appear at the end of the calibration (Fig. 23).