TT301
– Operation and Maintenance Instruction Manual
2.6
In a 2-wire connection, the voltage V2 is proportional to the RTD resistance plus the resistance of the
wires.
V2 = [RTD + 2
x
R]
x
I
R
V2
TRANSMITTER
2,1
3,4
R
RTD
I
Figure 2.3
– Two-Wire Connection
In order to avoid the resistance effect of the connection wires, it is recommended to use a 3-wire
connection (see Figure 2.4) or a 4-wire connection (see Figure 2.5).
In a 3-wire connection, terminal 3 is a high impedance input. Thus, no current flows through that wire
and no voltage drop is caused. The voltage V2-V1 is independent of the wire resistances since they will
be canceled out, and is directly proportional to the RTD resistance alone.
V2-V1 = [RTD + R]
x
I - R
x
I = RTD
x
I
R
V2
V1
TRANSMITTER
2,1
4
3
R
RTD
I
Figure 2.4
– Tree-Wire Connection
In a 4-wire connection, terminals 2 and 3 are high impedance inputs. Thus, no current flows through
those wires and no voltage drop is caused. The resistances of the other two wires are not interesting
since no measurement is done on them. Hence the voltage V2 is directly proportional to the RTD
resistance. (V2 = RTD
x
I).
Содержание TT301
Страница 2: ...www el scada ru...
Страница 8: ...TT301 Operation Maintenance Instruction Manual VIII...
Страница 16: ...TT301 Operation Maintenance Instruction Manual 1 8...
Страница 25: ...Operation 2 9 ALARM 1 ALARM 2 Figure 2 9 Typical Alarm Condition Display...
Страница 26: ...TT301 Operation and Maintenance Instruction Manual 2 10...
Страница 28: ...TT301 Operation Maintence Instruction Manual 3 2 Figure 3 1 Configurator Figure 3 2 CONF401 Screen...
Страница 47: ...Programming Using Local Adjustment 4 9 Configuration CONF Figure 4 7 Local Adjustment Configuration Tree...
Страница 58: ...TT301 Operation and Maintenance Instruction Manual 5 6 Figure 5 2 Four Possible Positions of the Display...
Страница 60: ...TT301 Operation and Maintenance Instruction Manual 5 8...
Страница 67: ...Certifications Information A 3 CSA NEMKO and DMT CEPEL...
Страница 69: ...Certifications Information A 5 CEPEL...
Страница 70: ...Operation and Maintenance Instruction Manual TT301 A 6 Control Drawing Factory Mutual FM...
Страница 71: ...Certifications Information A 7 NEMKO...
Страница 72: ...Operation and Maintenance Instruction Manual TT301 A 8 Canadian Standards Association CSA...
Страница 74: ...SRF Service Request Form B 2...