MeasuriNG aC voLtaGe
1. Connect the black measuring cord to the COM socket and the red cord to the V
Ω
F socket.
2. Set the dial to AC V.
3. You can now connect the cords to the circuit to be measured.
4. Power up the circuit to be measured, and you can read the measured AC voltage from the display.
Measurement range
resolution
Frequency
Precision (18 - 28 °C)
3,26 V
1 mV
40Hz-200Hz
± 0,8 % ± 3 dg.
32,6 V
10 mV
40Hz-400Hz
± 0,8 % ± 3 dg.
326 V
100 mV
40Hz-400Hz
± 0,8 % ± 3 dg.
750 V
1 V
40Hz-400Hz
± 1,2 % ± 3 dg.
Input impedance: 10 M
Ω
average sinusoidal value is displayed (AC rms)
MeasuriNG DC CurreNt
1. Connect the black measuring cord to the COM socket and the red cord to the A socket. If the value to be measured is larger than 300 mA or unknown, only
connect the red wire to the A socket.
2. Select the A function using the dial.
3. Now connect the instrument to the circuit (which is still switched off) in serial connection.
4. Power up the circuit to be measured, and the display will show the measured DC current. If you connected the measuring cords inversely, a minus sign
indicating negative polarity appears.
5. Once you have made sure that the current to be measured is below 300 mA, you can plug the red cord into the socket labeled mA after disconnecting the
power from the measured circuit. Then select the mA range using the dial. If you need to change the selected function or measurement range while taking
measurements, only do so after disconnecting the measuring cords out of precaution.
Measurement range
resolution
voltage drop
Precision (18 - 28 °C)
326 µA
0,1 µA
0,5 mV / µA
± 1,2 % ± 3 dg.
3,26 mA
1 µA
0,5 mV / µA
± 1,2 % ± 3 dg.
32,6 mA
10 µA
8,0 mV / mA
± 1,2 % ± 3 dg.
326 mA
0,1mA
8,0 mV / mA
± 1,2 % ± 3 dg.
10 A
10 mA
20 mV / A
± 2 % ± 5 dg.
Overload protection: F0.3 A fuse in the µA and mA ranges, F10 A fuse in the A range
MeasuriNG aC CurreNt
The procedure is identical to measuring DC current, but now you need to select the AC function by pressing button (3). The display will show the AC label.
Measurement range
resolution
Frequency
voltage drop
Precision (18 - 28 °C)
326 µA
0,1 µA
40Hz-400Hz
0,5 mV / µA
± 1,5% ± 3 dg.
3,26 mA
1
m
A
40Hz-400Hz
0,5 mV / µA
± 1,5% ± 3 dg.
32,6 mA
10
m
A
40Hz-400Hz
8,0 mV / mA
± 1,5% ± 3 dg.
326 mA
0,1 mA
40Hz-400Hz
8,0 mV / mA
± 1,5% ± 3 dg.
10 A
10 mA
40Hz-400Hz
20 mV / A
± 3,0% ± 5 dg.
Overload protection: F0.3 A fuse in the µA and mA ranges, F10 A fuse in the A range.
The average sinusoidal value is displayed (AC rms)
MeasuriNG resistaNCe
1. Connect the black measuring cord to the COM socket and the red cord to the V
Ω
F socket.
2. Select the
Ω
function using the dial.
3. If the resistor is not removed from the circuit, you must power off the circuit and discharge all capacitors.
4. You can now connect the cords to the part to be measured.
5. You can read the measured resistance from the display. The displayed value needs a few seconds to stabilize if you are measuring a resistor above 3.26 M
Ω
.
Measurement range
resolution
Precision (18 - 28 °C)
326
Ω
0,1
Ω
± 0,8% of rdg ± 3 dg.
3,26 k
Ω
1
Ω
± 0,8% of rdg ± 1 dg.
32,6 k
Ω
10
Ω
± 0,8% of rdg ± 1 dg.
326 k
Ω
100
Ω
± 0,8% of rdg ± 1 dg.
3,26 M
Ω
1 k
Ω
± 0,8% of rdg ± 1 dg.
32,6 M
Ω
10 k
Ω
± 1,2% of rdg ± 2 dg.
Maximum measuring voltage in an open circuit: 1.3 V
digital multimeter
GB
MY 68