background image

12. ACTIVATION ASSEMBLY REPLACEMENT

A. 

Tighten the retaining screw at the bottom right using Allen key

B.

 Slide up wall plate

C.

 Pull-out wall plate 

D.

 Disconnect Flushometer solenoid connector from the Sensor Assembly. Pull on end connector to prevent damage to wires.

E.

 Lift-up the Sensor Assembly Override Button to access the activation assembly.

F.

 Place the adjustment tool on the Activation Assembly and using a screwdriver, turn the assembly fully CLOCKWISE to make sure the valve is 

closed.

G.

 Remove the tool, reconnect the sensor assembly to the solenoid and then press the Override Button to relieve pressure.

H.

 Disconnect Flushometer solenoid connector from the Sensor Assembly (pull on end connector to prevent damage to wires), lift button up and 

push-in the activation assembly until fully seated.

I.

 Remove the orange safety ring by squeezing the pins together and pulling it out. 

NOTE. ATTACH SCREWS (INCLUDED) TO THREADED HOLES  

   ON ACTIVATION ASSEMBLY

J.

 Use attached screws to pull the activation assembly directly out, you may feel some tension. Be sure to pull directly out. 

Do not turn the   

HZZLTIS`[VH]VPK[\YUPUN[OL^H[LYÅV^IHJRVU

K.

 Install new Activation Assembly and press until fully seated.

L.

 Re-install the orange safety ring securely.  Safety ring will spin freely in valve body when properly installed.

M.

 Use the adjustment tool to slowly turn the activation assembly COUNTERCLOCKWISE to open the water flow.

N.

  Insert Flushometer Solenoid connector to the Sensor Assembly. 

+VUV[YLTV]LVYHKQ\Z[ZVSLUVPK

O.

 Activate Flushometer by pressing the Override Button.

P.

 Adjust water flow as needed (See Section 6).

Q.

 Put down the Sensor Assembly Override Button.

R. 

Install Wall Plate (See Section 7).

13. TROUBLESHOOTING

A. Sensor Flashes Continuously Only When User Steps Within Range.

          

i. Unit in Start-Up mode; no problem. This feature is active for the first ten (10) minutes of operation.

B. Valve Does Not Flush; Sensor Not Picking Up User.

        

  i. Range too short; increase the range (See Section 9).

C. Valve Does Not Flush; Sensor Picking Up Opposite Wall or Surface, or Only Flushes When Someone Walks By. Red Light Flashes  

    Continuously for First 10 Minutes Even with No One in Front of the Sensor.

         

 i. Range too long; shorten range (See Section 9).

+=HS]L+VLZ5V[-S\ZO,]LU(M[LY(KQ\Z[TLU[

         

 i. Ensure water supply to valve is turned on.

          ii. Activation Assembly sleeve is in closed position.  Turn counterclockwise to open water flow.

          iii. Batteries completely used up; replace batteries (See Section 10).

          iv. Problem with activation assembly; replace activation assembly (Section 12).

          v. Problem with Sensor Assembly; replace Sensor Assembly (See Section 3).

E. Red Light Blinks four (4) Times When User Steps Within Range (Battery Only).

          i. Batteries low; replace batteries (See Section 10).

F. Red Light Blinks four (4) Times When User Steps Within Range (Hardwire and Battery).

          i. Battery box not connected; connect battery box to sensor assembly (see section 10).

          ii. Batteries low; replace batteries (see section 10).

G. Valve Does Not Shut off.

          i. Metering bypass hole in piston is clogged. Remove the piston from the valve and wash under clean running water. Do not attempt to    

             remove filter screen. Replace piston if cleaning does not correct the problem (See Section 11).

          ii. Supply line water pressure has dropped and is not sufficient to close the valve.  Close Activation Assembly until pressure is restored.

          iii. Piston is damaged.  Replace with new proper gpf/Lpf piston. 

    

9

Содержание TRF 8156-1.28

Страница 1: ...8156 1 6 TRF 8156 1 28 TRF 8156 1 1 TRF 8196 0 5 TRF 8196 0 25 TRF 8196 0 125 PRIOR TO INSTALLATION TOOLS REQUIRED FOR INSTALLATION Smooth jawed wrench at least 2 Phillips screwdriver 5 64 hex wrench...

Страница 2: ...jaw wrench D Connect outlet pipe not porvided to flushometer Tighten securely with fixed jaw wrench NOTE DO NOT EXERT FORCE ON WALL BOX TO TIGHTEN THE FLUSHOMETER USE FIXED JAW WRENCH TO HOLD THE VALV...

Страница 3: ...f the pipe C Remove mud guard by removing the 2 50 mm long truss head screws and discard mud guard NOTE WALL PLATE OPENING MUST BE OF 5 6 X 5 6 142 X 142 MM D Install wall bracket onto flushometer usi...

Страница 4: ...racket i Align the 4 slots on the Sensor Assembly with the mounting pegs on the Wall Bracket ii Slide the Sensor Assembly all the way down For Hardwire connection use i Using a wire stripper strip the...

Страница 5: ...tart up routine is complete B After the start up routine is complete in the first ten 10 minutes of operation a visible Red LED flashes in the Sensor Window of the TruFlush Flushometer when a user is...

Страница 6: ...TTLE WATER FLOW MAY NOT SATISFY THE NEEDS OF THE FIXTURE PROPER ADJUSTMENT IS MADE WHEN PLUMBING FIXTURE IS CLEANSED AFTER EACH FLUSH WITHOUT SPLASHING WATER OUT FROM THE LIP AND A QUIET FLUSHING CYCL...

Страница 7: ...m Urinal Models 15 to 34 381 mm to 864 mm The Factory setting should be satisfactory for most installations If the range is too short i e not picking up users or too long i e picking up opposite wall...

Страница 8: ...assembly to the solenoid and then press the Override Button to relieve pressure H Disconnect Flushometer solenoid connector from the Sensor Assembly pull on end connector to prevent damage to wires li...

Страница 9: ...ee Section 6 Q Put down the Sensor Assembly Override Button R Install Wall Plate See Section 7 13 TROUBLESHOOTING A Sensor Flashes Continuously Only When User Steps Within Range i Unit in Start Up mod...

Страница 10: ...Valve Assembly 0 5 gpf 1 9 Lpf 3400108 TRF 8196 0 25 LP Sensor Valve Assembly 0 25 gpf 1 0 Lpf 3400109 TRF 8196 0 125 LP Sensor Valve Assembly 0 125 gpf 0 5 Lpf VALVE ASSEMBLY SELECTION GUIDE Filter S...

Страница 11: ...No 1 See Table Valve Assembly Less Plate Page 10 2 See Table Piston Assembly Page 10 3 See Table Activation Assembly Page 10 4a Wall Plate Assembly Sensor Closet 4b Wall Plate Assembly Sensor Urinal...

Отзывы: