background image

1. Sensor Flashes Continuously Only When User Steps Within 

Range.

A

. Unit in Start-Up mode; no problem. This feature is active for the first ten (10) 

minutes of operation.

2. Valve Does Not Flush; Sensor Not Picking Up User.

A

. Range too short; increase the range.

3. Valve Does Not Flush; Sensor Picking Up Opposite Wall or 

Surface, or Only Flushes When Someone Walks By. Light 
Flashes Continuously for First 10 Minutes Even with No One in 
Front of the Sensor.

A. Range too long; shorten range.

4. Valve Does Not Flush Even After Adjustment.

A

. Improper Range Adjustment; Follow Instructions and repeat steps.

B

. Batteries completely used up; replace batteries.

C

. Problem with Electronic Sensor Module; replace Electronic Sensor Module.

5. Light Blinks 4 Times When User Steps Within Range.

A. Batteries low; replace batteries.

6. Valve Does Not Shut Off.

A

. Bypass orifice in diaphragm is clogged with dirt or debris, or bypass is clogged 

by an invisible gelatinous film due to “over-treated” water. Remove flex tube 
diaphragm and wash under running water.

  Note: Size of Orifice in the Bypass is of utmost importance for 

the proper metering of water by the valve. DO NOT ENLARGE 
OR DAMAGE THIS ORIFICE.

 Replace flex tube diaphragm if cleaning 

does not correct the problem.

B

. Dirt or debris fouling stem or flex tube diaphragm. Remove flex tube 

diaphragm and wash under running water.

C

. O-ring on stem of flex tube diaphragm is damaged or worn. Replace O-ring if 

necessary.

D

. Problem with Electronic Sensor Module; replace Sensor Module. Try swapping 

out module with working unit to test module function.

E

. Are there audible clicks from the solenoid?

            

No

, possibly solenoid stuck inn open position, replace solenoid

           

 Yes

, Check solenoid properly screwed into inside cover, check to make    

         sure locking ring was secured properly.

F

. If module is working and solenoid clicking, check diaphragm for perforations in 

diaphragm disc; replace

7. Not Enough Water to Fixture.

A

. Wrong Flush Volume Regulator installed in Flex Tube Diaphragm Kit. Install the 

correct Regulator (see Section 7 of these instructions).

B

. Wrong G2/ECOS model installed; i.e., 1 gpf. urinal installed on closet fixture. 

Replace with proper G2/ECOS model, or refer to the G2/ECOS Conversion 
Guide to convert existing unit to the proper model.

C

. Enlarged Bypass in Diaphragm. Replace Flex Tube Diaphragm.

D

. Control Stop not adjusted properly. Readjust Control Stop.

E

. Inadequate volume or pressure at supply. Increase water pressure or supply 

(flow) to valve. Consult factory for assistance.

8. Too Much Water to Fixture.

A

. Wrong flush volume regulator installed in flex tube diaphragm kit. Install the 

correct regulator (see Step 7 of these instructions).

B

. Control Stop not adjusted properly. Readjust Control Stop.

C

. Wrong G2/ECOS model installed; i.e., 3 gpf. model installed on 1.0 or 1.5 gal. 

urinal fixture. Replace with proper G2/ECOS model, or refer to the G2/ECOS 

Conversion Guide to convert existing unit to the proper model.

D

. Dirt in diaphragm bypass. Clean under running water or replace flex tube 

diaphragm.

Note: The EBV-46-A Beam Deflector is no longer required or 
available for the G2/ECOS.

7

LAWS AND REGULATIONS PROHIBIT THE USE OF 

HIGHER FLUSHING VOLUMES THAN 

LISTED ON FIXTURE OR FLUSHOMETER

!!! IMPORTANT !!!

THE CONTROL STOP SHOULD NEVER BE OPENED 

TO THE POINT WHERE THE FLOW FROM THE VALVE 

EXCEEDS THE FLOW CAPABILITY OF THE FIXTURE. IN 

THE EVENT OF A VALVE FAILURE, THE FIXTURE MUST BE 

ABLE TO ACCOMMODATE A CONTINUOUS FLOW FROM 

THE VALVE.

!!! IMPORTANT !!!

WITH THE EXCEPTION OF CONTROL STOP INLET, DO 

NOT USE PIPE SEALANT OR PLUMBING GREASE ON ANY 

VALVE COMPONENT OR COUPLING!

!!! IMPORTANT !!!

PROTECT THE CHROME OR SPECIAL FINISH OF SLOAN 

FLUSHOMETERS — DO NOT USE TOOTHED TOOLS TO 

INSTALL OR SERVICE THESE VALVES. USE A SLOAN 

A-50 SUPER-WRENCH™, SLOAN A-109 PLIER WRENCH 

OR SMOOTH JAWED SPUD WRENCH TO SECURE ALL 

COUPLINGS. ALSO SEE “CARE AND CLEANING” SECTION 

!!! IMPORTANT !!!

THIS PRODUCT CONTAINS MECHANICAL AND/OR 

ELECTRICAL COMPONENTS THAT ARE SUBJECT TO 

NORMAL WEAR. THESE COMPONENTS SHOULD BE 

CHECKED ON A REGULAR BASIS AND REPLACED AS 

NEEDED TO MAINTAIN THE VALVE’S PERFORMANCE.

!!! IMPORTANT !!!

THE STRAP WRENCH PROVIDED WITH SLOAN ECOS

®

 

ELECTRONIC IS A CONVENIENCE TOOL AND IS NOT TO 

BE USED TO REMOVE OR INSTALL THE FLUSHOMETER 

COUPLINGS. USE STRAP WRENCH ONLY TO INSTALL 

SLOAN ECOS

®

 ELECTRONIC LOCKING RING.

!!! IMPORTANT !!!

RECOMMENDED WALL PLATE LOCATIONS

CARE AND CLEANING

• Centered over Flushometer   

• On stall door

DO NOT use abrasive or chemical cleaners to clean flushometers as they may 

dull the luster and attack the chrome or special decorative finishes. Use ONLY 

soap and water, then wipe dry with clean cloth or towel.
While cleaning the bathroom tile, the Flushometer should be protected from 

any splattering of cleaner. Acids and cleaning fluids can discolor or remove 

chrome plating.

When required, replace batteries with four (4) Alkaline 

AA-Size Batteries.

Note: Water does not have to be turned off to 
replace Batteries.

Loosen the two (2) Screws on top of unit. Remove the 

complete Cover Assembly. Lift the Sensor Module from 

its Plate. Unplug the Electrical Connector from Battery 

Compartment Cover. Loosen the Retaining Screw 

on Battery Compartment Cover and remove Battery 

Compartment Cover. Install four (4) Alkaline AA-Size 

Batteries exactly as illustrated.

BATTERY 

COMPARTMENT 

COVER

ELECTRICAL CONNECTOR RECEPTACLE

SENSOR 

MODULE

RETAINING 

SCREW

16 - TROUBLESHOOTING

15- BATTERY REPLACEMENT

COVER 

ASSEMBLY

SENSOR 

MODULE

PLATE

Install Battery Compartment Cover and secure with Retaining 

Screw. Make certain that Battery Compartment Cover is fully 

compressed against Gasket to provide a seal; Do Not overtighten. 

Plug the Electrical Connector into the Battery Compartment Cover. 

Reinstall the Sensor Module onto the Plate. Reinstall the complete 

Cover Assembly onto the Plate. Tighten the two (2) Screws on top 

of the unit.

Содержание ECOS HET 8100 Series

Страница 1: ...WHEN INSTALLING A FLUSHOMETER IT IS IMPORTANT THAT THE FLUSH MODEL MATCHES THE REQUIREMENTS OF THE PLUMBING FIXTURE FLUSH ALL WATER LINES PRIOR TO MAKING CONNECTIONS The Sloan ECOS is designed to oper...

Страница 2: ...allation Model 8110 8111 Reference for RESS C Retrofit Model 8115 8116 valves are designed for installations where the water supply is roughed in 24 27 610 mm 686 mm above the top of the water closet...

Страница 3: ...Sweat solder the Adapter to pipe WITH THE EXCEPTION OF CONTROL STOP INLET DO NOT USE PIPE SEALANT OR PLUMBING GREASE ON ANY VALVE COMPONENT OR COUPLING IMPORTANT BAK CHEK CONTROL STOP COVER TUBE IRON...

Страница 4: ...and Spud Flange over Vacuum Breaker Tube B Insert Tube into Fixture Spud C Hand tighten Spud Coupling onto Fixture Spud G 44 FRICTION RING C L SUPPLY ADJUSTABLE TAILPIECE C L FIXTURE CONTROL STOP O RI...

Страница 5: ...fitting the Optima Plus onto a Zurn valve body a special Locking Ring must be used identified by a machined groove around the ring Order the Optima Plus with the Z variation to receive the unit suppli...

Страница 6: ...w may not satisfy the needs of the fixture Proper adjustment is made when plumbing fixture is cleansed after each flush without splashing water out from the lip AND a quiet flushing cycle is achieved...

Страница 7: ...xture Replace with proper G2 ECOS model or refer to the G2 ECOS Conversion Guide to convert existing unit to the proper model D Dirt in diaphragm bypass Clean under running water or replace flex tube...

Страница 8: ...ete Sloan ECOS Valves Only 17 H 633 AA Sweat Solder Kit 18 H 700 A Bak Chek Control Stop 19 H 1010 A Vandal Resistant Stop Cap 20 EBV 36 A Valve Body 21A V 600 AA 1 38 mm x 10 254 mm Vacuum Breaker Mo...

Отзывы: