OPERATION
1.
Use the following procedure to test the system before each use.
a.
Press Battery-Test button. If alarm is weak or fails to sound,
replace 9-volt battery and press button again. If alarm is still weak or
fails to sound, discontinue use of alarm.
b.
Seat patient and secure belt by placing the magnet (located
under red loop) over metal strip. (Illustration 3) Battery indicator light
should flash. If light fails to flash, replace battery and make certain that
the telephone-style plug is securely connected to the alarm unit. Retest.
If battery indicator light still fails to flash, discontinue use of system.
c.
Open magnetic closure. Alarm will sound. Secure closure.
Alarm will silence itself. It is now in the monitoring mode. If alarm fails
to sound when magnetic
closure is opened, check
plug connection and/or
replace battery. Retest. If
alarm again fails to
sound, discontinue use
of system.
2.
Select high or low vol-
ume.
3.
Alarm is activated
when patient opens the
magnetic closure. Press
the Auto-Reset button to
silence the alarm. The
alarm is now in stand-by
mode. Do not secure the magnetic closure when the system is not
being used to monitor a patient. Securing the closure will cause the
battery indicator light to flash and shorten battery life.
4.
The alarm will automatically go into the monitoring mode when the
magnetic closure is secured.
There is no on/off switch.
5.
After the closure is secured, adjust the belt to a proper fit around
patient’s hips.The wheelchair user may easily release the belt by plac-
ing his/her hand through the red loop and pulling up on the end.
with E-Z Release Velcro
®
Belt Alarm
29 Wells Avenue
914-963-2040
Yonkers, NY 10701
800-431-2972
Model # Description
Guarantee Unit
909393
E-Z Release Velcro
®
Belt Alarm One Year
Each
INSTALLATION
1.
Remove the rearmost screws
from the left and right sides of
the wheelchair seat. Insert the
screws through the grommets
on the ends of the two belt sec-
tions. Refasten screws to the
wheelchair. (Illustration 1)
2.
Place the telephone-style
connector cord through the
space between the wheelchair
seat and backrest. Make certain
that it is away from wheels and
other moving parts.
3.
Insert telephone-style plug into
Modular Receptacle on right side
of alarm. (Illustration 2) Alarm will
sound. Press Auto-Reset to
silence alarm.
Advanced design for nursing and therapy.
TM
Alarm system made in China. Instructions printed in USA.
Visit us online
at Skil-Care.com
1
2
3
MultiPro/ChairPro E-Z Releas.Qk 4/3/06 9:38 AM Page 1
H
WHEELCHAIR INSTALLATION (Loop) cont.
5. Then place webbing over ring “A” and under ring “B” (Fig. 3)
6. Pull end until taut. Place arm back on wheelchair and repeat again
for alternate side of wheelchair (Fig. 4)
After both belts are attached to the
wheelchair, place the belt around the
resident and adjust the belt so that it is
snug but not uncomrfortably tight.
GERI-CHAIR/UTILITY CHAIR INSTALLATION (Loop)
1. Unthread the loop end of the belt.
2. Wrap loop end of the belt around the
chair leg (Fig. 1).
3. Thread the loop end of the belt
under both rectangular rings (Fig. 2).
4. Then place webbing over ring “A” and
under ring “B” (Fig. 3). Do not tighten
loop until the belt is adjusted.
5. Thread end of buckle through the seat
and place on seat.
6. Repeat process for other side of chair
7. Adjust length of belt so that it is snug
but not too tight.
Note: There are adjustments with the
tri-glide on he belt and to a lesser degree,
adjustment with the loop end.
OPERATION
1. Remove battery compartment cover on left side of alarm unit and install two “AA” batteries.
2. Press Battery Test button. If alarm doesn’t sound, check battery connection and retest. If alarm fails
to sound, replace battery.
3. Insert mini-phone jack into Mini-Phone Plug receptacle on right side of alarm. (Illustration G) Alarm
will sound. Press Auto-Reset to silence alarm.
4. Before using, test unit by placing magnet (located under red loop) over
metal strip. (Illustration H) Check alarm to make certain that LED is glowing. Separate the magnet
from the metal strip. Alarm should sound. Press Reset button to silence alarm. If alarm fails to sound,
make certain that spade connectors and phone jack are securely in place.
5. Select High or Low alarm volume using switch on right of alarm.
6. Important: If the alarm is activated, it may be silenced by pressing the Auto-Reset button.
The alarm is now in stand-by mode. Do not secure buckle when unit is not being used.
Doing so will cause the LED to glow and shorten battery life. Alarm will
automatically reset itself when the metal tab is inserted into the buckle. There is no
On/Off switch.
7. Place belt around user’s waist and secure by placing tab into buckle. Adjust belt to
proper fit.
Use the slider buckle to adjust the length of belt. If the belt is too long or short see
step 4 under “Wheelchair Installation (Grommet)”.
SAFETY REMINDERS
1. Check battery before each use. If alarm is weak or fails to sound, replace two “AA” batteries.
2. When unit is in use, check LED regularly. If it is not glowing, replace two “AA” batteries.
GUARANTEE
The alarm unit, belt, and buckle are replacement guaranteed against defects in materials and workmanship under conditions
of normal use for a period of one year from date of purchase. Please return defective part to dealer from whom the system was