SKF TKSA 20

b)  Aim the laser lines so that they hit in the centre of the target of the opposite 

measuring unit (fig. 11).

B

0

B

0

Fig. 11. Hit the target

A Laser line

c)  For coarse adjustment release the measuring unit by unlocking the knob on the side 

of the unit (fig. 12). This allows the measuring unit to slide up and down the rod at the 

same time as it can swivel freely. For the fine adjustment in height use the adjustment 

wheels on the measuring units.

0

2

1

Fig. 12. Adjustment mechnism

A

B

C

Vertical positioning of measuring unit

Horizontal rotation of measuring unit

Vertical fine adjustment of laser

C

Содержание TKSA 20

Страница 1: ...Instructions for use SKF Shaft Alignment Tool TKSA 20...

Страница 2: ......

Страница 3: ...lts 17 3 8 1 Measured misalignment 17 3 8 2 Vertical alignment 17 3 8 3 Horizontal alignment 19 3 9 Verify alignment 20 3 10 Soft foot 21 4 Alignment report 22 5 Advanced use 24 5 1 Limited rotation 2...

Страница 4: ...E 2004 108 EC as outlined in the harmonized norm for Emission EN 61000 6 3 2007 Immunity EN 61000 6 2 2005 EN 61000 4 2 3 Directive RoHS 2002 95 EC The laser is classified in accordance with the EN 60...

Страница 5: ...t power below 1 mW Still never stare directly into the laser transmitter Calibrate the equipment regularly Never aim the laser line into someone s eyes Opening the housing of the measuring unit may re...

Страница 6: ...ser lines to deflect from their initial relative position The measurements from the two positioning detectors automatically enter the logic circuitry inside the display unit which calculates the misal...

Страница 7: ...ment procedure we use the analogy of a clock as viewed from behind the Movable machine The position with the measuring units in an upright position is defined as 12 o clock while 90 left or right is d...

Страница 8: ...isplay unit 2 measuring units with spirit levels 2 mechanical shaft fixtures 2 locking chains Measuring tape Quick start guide Certificate of calibration CD ROM including Instructions for use Quick st...

Страница 9: ...r for measuring unit on Movable machine 3 LCD Display 4 ON OFF button 5 Increase button 6 Next button 7 Decrease button 8 Previous button 9 Machine dimensions A B and C Measured values S and M 10Rear...

Страница 10: ...fixture with measuring unit 1 2 3 4 5 Laser emission Laser warning signal Laser detector Vertical fine adjustment Spirit level 6 7 8 9 10 Release tightening knob Connection rod Chain fixation screw Lo...

Страница 11: ...mm 3 4 x 3 1 x 1 5 in 210 gram 7 3 oz Display unit Housing material Display type Battery type Operating time Auto power off Displayed resolution Dimensions Weight ABS plastic LCD 35 x 48 mm 1 4 x 1 9...

Страница 12: ...so called soft foot The procedure for this operation is described in chapter 3 10 3 3 Attachment of measuring units a Use the fixtures to attach the measuring units to the shafts Make sure that the M...

Страница 13: ...easuring units 3 4 Switch on Switch on the display unit by pressing the ON OFF button You will now be prompted to enter the machine dimensions as per chapter 3 6 If no button is pressed within 60 minu...

Страница 14: ...asuring unit by unlocking the knob on the side of the unit fig 12 This allows the measuring unit to slide up and down the rod at the same time as it can swivel freely For the fine adjustment in height...

Страница 15: ...units to the 3 o clock position when the lines will hit outside the detector areas Adjust the lines to the position half way between the detector centre and the actual position by means of the adjustm...

Страница 16: ...he Movable machine C The distance between the front feet and the rear feet of the Movable machine 0 2 1 Fig 14 Machine dimensions a Measure the A B and C distances The default values for these three d...

Страница 17: ...bed earlier chapter 1 3 we use the analogy of a clock to describe the different positions Fig 16 Display guides you to the 9 o clock position a Adjust the measuring units to the 9 o clock position wit...

Страница 18: ...e parallel off set of the two centre lines in the measurement plane These two values are the coupling values in the measurement plane Feet values The values F1 and F2 on the display indicate the relat...

Страница 19: ...3 Horizontal alignment b If the measured coupling values are higher than the acceptable tolerances check the recommended corrections of the feet The F1 and F2 values on the display indicate the relat...

Страница 20: ...nment should be within the manufacturer s specified tolerances In case such tolerances are missing table 1 can again be used as a general recommendation a If the measured coupling values are within th...

Страница 21: ...to the right A positive value means that the feet have to be moved to the left fig 26 Observe the coupling and feet values live adjustment while moving the movable machine sideways and compare them wi...

Страница 22: ...M mm F1 F2 Fig 28 Soft foot display 6 Loosen one of the bolts and monitor the change of the displayed values Monitor the changes of the F1 value for a front foot and of the F2 value for a rear foot I...

Страница 23: ...ximum acceptable off set of centre lines see table 1 Ch 3 8 2 i Selection of metric or imperial dimensions j Machine configuration distances A B and C k Soft foot correction done l Vertical alignment...

Страница 24: ...23 SKF TKSA 20 B 5 BWPUc P XV T c aT_ ac P Q R S T U V W X Y Y Y Z Z Z Z _ i a b c d e f g h i i i i i X...

Страница 25: ...asurement 3 You can now complete the alignment following the instructions sequence as per chapter 3 8 5 2 Trouble shooting 5 2 1 The system does not switch on a Check that batteries are inserted in th...

Страница 26: ...ents in the machinery c Check the status of foundation base plate bolts and existing shims 6 Maintenance 6 1 Handle with care The measuring units are equipped with sensitive electronic and optical par...

Страница 27: ...units Movable and Stationary TKSA and TMEA 2 system Locking chains set 500 mm tightening tool Extension chains set 1020 mm 1 chain fixture complete Set with 3 pairs of connection rods short 150 mm st...

Страница 28: ...is granted Every care has been taken to ensure the accuracy of the information contained in this publication but no liability can be accepted for any loss or damage whether direct indirect or consequ...

Отзывы: