26
SKF TKGT 1
3.5.4 Recommendations
Once the test has been done for a fresh unused grease sample, record the value for later.
Only the used sample test will need to be performed for further analysis.
Make sure that the fresh sample and used samples are heated within the same
temperature range (+/- 5 °C will not considerably inluence the result).
Perform the test at room temperature 15 °C to 30 °C (59 °F to 86 °F). If temperature is
below or above, the heater might not reach the testing temperature. The USB heater might
not reach testing temperature if test is performed in a windy environment.
Severe contamination can, in certain cases block oil bleeding and result will show a
consequent difference, while the grease still seems to be oily. This might happen if the
sample taken is a mix of used grease and grease added during a relubrication interval.