Preparation before test:
•
Print the calibrated measuring scale with right printer settings. Do not scale to page!
To check if the print out is the correct one, the outside diameter of the housing should
be the same as the circle drawn on the calibrated measuring scale.
•
Grease samples should be at temperature between 15 °C and 30 °C (59 °F and 86 °F).
•
Wristwatch or clock to measure 15 seconds (not provided).
NLGI Grade
000
00
0
1
2
3
4
5
13
12
10
11
B
3.4.3 Test procedure
1. Ensure all components are clean
2. Wear gloves
3. Place weight in the housing
4. Place one glass plate on top of the
weight
5. Place mask on top of the glass plate
6. Take the grease sample to analyze from the
sample container using a disposable spatula
7. Apply the grease in the mask. Make sure that hole is full with grease and wipe off excess on the top.