21
Calibration
7. When calibration is completed, the pump will go to HOLD. If the
calibration was successful, the pump LCD will revert to displaying pump
run time as 0. If there was failure during the calibration process, an error
code of E4[x] will appear (
see CalChek Error Chart on page 22
).
Note:
To remove a CalChek error code from the LCD, press
Ò
.
8. Allow the pump to go to SLEEP mode to write calibration data to pump
memory.
Data does not write to pump memory until pump has gone into
SLEEP mode.
CalChek Full Calibration Data
Full calibration data can be viewed and printed by going to the DataTrac
2000 Pump Manager window and clicking on the View menu. Choose
Calibration Info. This will display calibration results, pump serial number,
and date of the last full calibration. A bu
tt
on allows this data to be printed. The
printed report contains pump version, date printed, and a validation code.
CalChek Calibration Data Veri
fi
cation in DataTrac 2000 Software
This feature requires pump version 2.62 or higher and DataTrac 2000
So
ft
ware version 3.62 or higher.
To ensure that printed data has not been tampered with, pull down the Tools
menu and choose Con
fi
rm Validation Code. Enter the data from the printed
report, including the validation code. DataTrac 2000 will indicate whether the
information is completely valid or if a parameter has been changed.
Clearing the history will not clear full calibration data. This data can
only be cleared by performing another full calibration or by obtaining
more than 36 pump history records.
When entering data to con
fi
rm the validation number, enter the date
in the following format: mmm dd, yyyy (e.g., Aug 18, 2008)