![SKC AirChek2000 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/skc/airchek2000/airchek2000_operating-instructions-manual_1286851017.webp)
14
Setting a Delayed Start
The Delayed Start feature is available in pump version 2.59 or higher.
When se
tt
ing the pump manually for sampling to begin within a 12-hour
timeframe, follow this procedure:
1. With the pump running, press [
ST
] to place the pump in HOLD and
enter
Ò
ST
Ò
within 10 seconds.
You are in User Interface Level Two.
2. Press
Ò
until
S
Г
appears. Press
S
or
T
to set a sample time. This is
required to set a delayed start.
3. Press
Ò
until the display reaches the 12-hr/24-hr clock. Press
S
or
T
until the display shows a
fl
ashing dELA (delayed start).
4. Press
Ò
. Time will display with hours
fl
ashing. Press
S
to increase or
T
to decrease the hour. Press
Ò
to move to minutes. Press
S
to increase
or
T
to decrease the minutes.
The time entered here will be the next occurrence of this time. There
is no a.m. or p.m. designation.
5. Press
Ò
until End appears.
6. Press
[
ST
].
7. The PROG icon and a
fl
ashing HOLD will appear in the upper le
ft
corner of the display. The pump is now set for delayed start.
If se
tt
ing a delayed start using DataTrac 2000 So
ft
ware, refer to the
DataTrac 2000 Operating Instructions (included on the so
ft
ware CD).
A programmed delayed start can be viewed in the DataTrac 2000
Scheduler.
Deleting a Delayed Start
This is the same method used to delete a DataTrac 2000 program from the
pump (
see Deleting a DataTrac 2000 Program on page 15
).
1. With the pump running (will only run in short bursts), press [
ST
] to
place the pump in HOLD and enter
Ò
ST
Ò
within 10 seconds.
You are
in User Interface Level Two.
2. Press
Ò
to scroll to the
fl
ashing OFF indicator and press [
ST
].
3. Press
Ò
until End displays.
4. Press
[
ST
]. The PROG icon should be gone. If not, repeat process until
it disappears.
Setup