Operating the Flite4 Pump
On the ‘SELECT Hold’ screen cycle to the ‘Hold’ option
and press to stop the pump running and display the
Hold menu screen, which also shows the current elapsed
runtime.
SELECT
Hold Continue
Æ
Reset 001:00:06
Continue Flowset
To check and adjust the pump flow rate, on the Hold menu
screen cycle to the ‘Flowset’ option and press . The
pump will run and display the ‘SET FLOW’ screen.
Note:
The elapsed runtime does not increment whilst
checking and adjusting the flow rate.
Reset 001:00:06
Continue Flowset
Æ
SET FLOW Back
>>>>>>__________
Adjust the pump flow rate using a suitable flow calibrator
and press to save the new flow rate. The pump will stop
running and return to the Hold menu screen.
SET FLOW Back
>>>>>>>_________
Æ
Reset 001:00:06
Continue Flowset
To continue the sample run, on the Hold menu screen
cycle to the ‘Continue’ option and press . The pump will
resume running and the elapsed runtime screen will be
displayed and continue to increment.
Reset 001:00:06
Continue Flowset
Æ
ELAPSED RUN TIME
001:00:09