Programming the Flite4 Pump
1) Setting the Flow Rate
Fit a new filter into the sampling head and connect the sampling head outlet to the pump inlet hosetail
using a length of flexible tubing. Connect the sampling head inlet to the outlet of a flow calibration device,
such as a chek-mate electronic flowmeter or rotameter.
Cycle to the ‘Setup’ option in the ‘SELECT’ main menu
screen, and press to display the ‘Setup’ menu screen.
SELECT Sleep
Run Setup Review
Æ
Set Clear Back
Flow RunTime
On the ‘Setup’ menu screen cycle to the ‘Flow’ option and
press . The pump will start to run and the ‘SET FLOW’
screen will be displayed.
Set Clear Back
Flow RunTime
Æ
SET FLOW Back
>_______________
Press
to increase the pump flow rate. More arrow
symbols are displayed on the LCD screen to indicate the
increased flow rate.
Note:
Pressing and holding will rapidly increase the
pump flow rate.
When the maximum possible pump flow rate is reached,
this is indicated on the LCD screen as shown.
Æ
SET FLOW Back
>>>>>>>>>>>>____
SET FLOW Back
>>>>>>>>>>>>>Max