![SKC 224-44XRM Скачать руководство пользователя страница 3](http://html.mh-extra.com/html/skc/224-44xrm/224-44xrm_operating-instructions-manual_1286834003.webp)
2
Performance Pro
fi
le
Operating Range
...................1000 to 3000 ml/min
(5 to 500 ml/min requires adjustable low
fl
ow holder)
Weight
....................................34 oz (964 gm)
Dimensions
...........................5.1 x 4.7 x 1.9 in (13 x 11.9 x 4.8 cm)
Compensation Range
...........1000 ml/min at 40 inches water back pressure
2000 ml/min at 40 inches water back pressure
3000 ml/min at 22 inches water back pressure
Flow Control
..........................Holds constant
fl
ow to ± 5% of the set point
Run Time
...............................8 hrs minimum at 3000 ml/min and 20 inches water back pressure;
dependent on media used
Flow Indicator
.......................Built-in
fl
ow indicator with 250-ml division; scale marked at 1, 2, 3,
4, and 5 L/min
Power Supply
........................MSHA approved when used with rechargeable NiCad battery pack,
2.0 Ah capacity, 4.8 V, Cat. No. P22434 only. Recharge battery pack
in fresh air only.
Charging Time
......................16 hrs with approved charger
(varies with capacity and
level of discharge)
Operating Temperature
........32 to 113 F (0 to 45 C)
Storage Temperature
...........-40 to 113 F (-40 to 45 C)
Charging Temperature
.........41 to 113 F (5 to 45 C)
Operating Humidity
..............0 to 95% Relative
Protect sample pump from exposure to water and weather.
Multiple-tube Sampling
........Built-in constant pressure regulator allows user to take up to four
simultaneous samples at different
fl
ow rates up to 500 ml/min each
using an adjustable low
fl
ow holder
RFI-Shielding
Performance
..........................Complies with requirements of EN 55022, FCC Part 15 Class B,
EN 50082-1; frequency range of the radiated susceptibility test was
27 MHz to 1000 MHz. CE marked.
Intrinsic Safety
......................MSHA-approved to 30 CFR, section 18; approval no. 2G-4036-0
Tested for use in methane-air mixtures only
Intrinsic safety and other approvals may be void if SKC pumps are not
repaired by SKC or authorized SKC repair centers.
Replacing parts with non-SKC parts voids intrinsic safety approvals
and any warranty.
Содержание 224-44XRM
Страница 15: ...14 Figure 1 224 44XRM Sample Pump See parts listing on page 13 ...
Страница 19: ...18 MSHA Approval Certificate ...