Pump Inlet Filter Maintenance
The SKC sampler is
fi
tted with a
fi
lter/trap inside a clear plastic intake port housing.
This prevents particles from being drawn into the pump mechanism. The
fi
lter should
be checked visually to assure that it does not become clogged. If maintenance is
necessary, follow this procedure:
1. Clean dust and debris from around the
fi
lter housing.
2. Remove the four screws (Figure 1, #7) and the front
fi
lter housing.
3. Remove and discard the
fi
lter membrane (Figure 1, #9).
4. Remove O-ring (Figure 1, #8).
5. Clean
the
fi
lter housing.
6. Insert O-ring* and a new
fi
lter membrane (Cat. No. P2240901).
7. Reattach the front
fi
lter housing and cross-tighten the four screws.
* Replace with new O-ring only as needed.
Pump Service
Pumps under warranty should be sent to SKC Inc. for servicing (