background image

-  For continue protection against fire, replace fuse only with the  

  specified voltage and current ratings: F 250mA/600V, 

  F 10A/600V(quick acting)

-  Do not use abrasives or solvents on the meter, use a damp cloth  

  and mild detergent only.

-  Always set the power switch to the OFF position when the meter  

  is not use.

-  If the meter is to be stored for a long period of time, the batteries  

  should be removed to prevent damage to the unit.

DESCRIPTION

-  This meter is a portable professional measuring instrument with  

  handsome LCD easilyreading.

-  Single operation of a transform switch makes measurement con 

  venient. Overload protection and low battery indication are  

  provided, this meter is ideal for use in the fields, workshop, school,  

  hobby and home applications.

-  This meter is with the fuctions of ‘data hold’.

NAME OF COMPONENTS

1. LCD display

2. ‘Hold’ key

3. Rotary switch

4. Terminals

SPECIFICATIONS

-   Environment conditions:

  600V CAT.III 

  Pollution degree: 2. Altitude < 2000 m.

-  Operating temperature: 

 0~40°C (32°F - 104°F), (<80% RH,non-condensing)

-  Storage temperature: 

  -10~50 OC(14°F - 122°F), (<70% RV, battery removed)

-   Temperature Coefficient:

  0,1×(specified accuracy) / OC (<18°C of >28°C)

-   MAX. Voltage between terminals and earth ground: 

  600V AC rms of 600V DC

-  Fuse Protection: mA: F 250mA/600V ø 6.3×30, 

  A: F 10A/600V ø 6.3×30.

-  Display: LCD, 1999counts, updates 2-3/sec.

-  Over Range indication: LCD will display “1”.

-  Low battery indication: 

  The “

“ is displayed on the LCD.

-  Polarity indication: “

“ displayed automatically.

-  Power source: 9V    

-   Battery type: 6F22 9V

-  Dimensions: 158(L)×74(W)×37(H) mm.

-  Weight: 220 g. Approx. (battery included).

MEASUREMENT SPECIFICATIONS

Accuracy: ±(% r number of digits) at 18°C - 28°C 

(64°F - 82°F) with relative humidity to 80%.

(Accuracy is specified for a period one year after calibration).

DC Voltage

Range  Resolution Accuracy
0.2V 0.1mV   
2V 1mV  ±0.5%±1 

20V 0.01V   
200V 0.1V 

 

600V 1V 

±0.8%±2 

-  Input impedance: 10M

Ω

-  Max. input voltage: 200mV range: 250VDC or AC rms 
  2V - 600V ranges: 600VDC or AC rms.

AC VOLTAGE

Rance  

Resolution  Accuracy

2V 1mV  ±0.8%±3 

20V 10mV   
200V 0.01V   
600V 1V 

±1.2%±3

Input impedance:10M

Ω

Max. iinput: 200mV range: 250VDC or AC rms
           2V - 600V ranges: 600VDC or AC rms
Frequency range: 40Hz-1000Hz
Response: Average, calibrated in rms of sine wave.

Skandia Digital Multimete 1088025

• See the instructions before use.

SAFETY INFORMATION

 Warning

To ensure safe operation and in order to exploit to the full the 

functionality of the meter, please follow the directions in this  

section carefully.

This multimeter has been designed according to IEC-61010  

concerning electronic measuring instruments with an overvoltage 

category CAT III 600V and pollution 2. Only to use for measurements 

in a 1-phase elektrical 600V installation after the group distribution 

frame.! Also to use for measurements of electric appliances with a 

plug and lighting connected on this installation after the group  

distribution frame.! Suitable for measurements of dc voltage and  

AC voltage till maximum 600V.

Follow all safety and operating instructions to ensure that the meter 

is used safely and is kept in good operating condition.

With proper use and care, the digital meter will give you years 

of satisfactory service.

PRELIMINARY

-   When using the meter, the user must observe all normal safety  

  rules concerning:  

  Protection against the dangers of electrical current.

  Protection of the meter against misuse.

-  When the meter is delivered, check that it has not been  

  damaged in transit.

-  When poor condition under harsh preservation or shipping  

  conditions caused, inspect and confirm this meter without  

 delay. 

-  Test leads must be in good condition. Before using verify  

  that the insulation on test leads is not dameged and/or the  

  leads wire is not exposed.

-  Full compliance with safety standards can be guaranteed  

  only if used with test leads supplied. If necessary they must  

  be replaced with the same model or same electric ratings.

DURING USE

-  Before using, you must select the right input jack, function and  

 range.

-  Never exceed the protection limit values indicated in specifications  

  for each range of measurement.

-  When the value scale to be measured is unknown beforehand, set  

  the range selector at the highest position.

-  Do not measure voltage if the voltage on the termanals exeeds  

  600V above earth ground.

-  Always be careful when working with voltages above 60V DC  

  or 30V AC rms, keep fingers behind the probe barriers while  

 measuring.

-  Before rotating the transform switch to change functions and  

  ranges, disconnect test leads from the circuit under test.

-  Never perform resistance, temperature, transistor, diode and  

  continuity measurements on live circuits.

-  Never use the meter under the condition of the explosive air,  

  stream or dirt.

-  If any faults or abnormalities are observed, the meter can  

  not be used any more and it has to be checked out.

-  Never use the meter unless the rear case is in place and  

  fastened fully.

-  Please do not store or use meter in areas exposed to direct  

  sunlight, high temperature, humidity or condensation.

 SYMBOL

  

Caution:

 refer to the instruction manual. Incorrect 

   

use may result in damage tothe device or its 

  

components.

 

  

Earth

    

This instrument has double insulation.

 

   

Fuse: F 250mA/600V, F 10A/600V.

MAINTENANCE

-  please do not attempt to adjust or repair the meter by removing  

  the rear case while voltage is being applied. A technician who fully  

  understands danger involved should only carry out such actions.

-  Before opening the battery cover or case of the meter, always  

  disconnect test leads from all tested circuits.

-  To avoid the wrong reading causingelectricity attack, when the  

  meter displays “ 

 ”, you must change the battery.  

INSTRUCTION

DC Current

Rance  Resolution Accuracy
200uA 0.1uA 

 

2mA 1uA 

±0.8%±1 

20mA 10uA 

 

200mA 100uA 

±1.2%±1 

10A 10mA  ±2.0%±5 
Overload protection: F 250mA/600V, F 10A/600V fuse. 
Max. input current: mA: 250mA, 10A Jack: 10A

AC Current

Rance  Resolution Accuracy
2mA 1uA 

±1,2%±3

20mA 10uA 
200mA 0.1mA 

±2,0%±3

10A 10mA  ±3,0%±7

Overload protection: F 250 mA/600V, F 10A/600V fuse. 

Max. input current: mA: 250mA dc or 250mA ac rms 

10A: 10A continuous, 10A 15 sec. Max.

Frequency Range: 40Hz-1kHz 

Indication: Average (rms of sine wave) 

Resistance

Rance  Resolution Accuracy
200

Ω

 0.1

Ω

 ±0,8%±3 

2K

Ω

 1

Ω

 

20K

Ω

 10

Ω

 ±0,8%±1   

200K

Ω

 100

Ω

  

2M

Ω

 1k

Ω

  

20M

Ω

 10k

Ω

 ±1,0%±2 

Overload protection: 250 DC or 250  AC rms.

Diode test

 Display: read approximate forward voltage of diode

 

Forward DC Current: approx. 1mA

      Reversed CD Voltage: approx. ca. 2,8V
      Overload protection: 250V DC or 250V AC rms

Audible continuity test

    Built-in buzzer sounds if resistance is less than approx. 50

Ω

.

      Open circuit voltage: approx. 2,8V
      Overload protection: 250V DC or 250V AC rms.

Battery Test

 

Rance 

Accuracy 

Discharge current

1,5V 

±0,8%±1 100mA

9V  

6mA

Overload protection: 1.5V: F 250mA/600V fuse 
9V: 250V dc or 250V ac rms.

OPERATING INSTRUCTION

Voltage measurement

-   Set rotary switch to the desired V 

 or V~ range.

-  Connect the black and red test leads to the COM and V terminals  

 respectively.

-  Connect the test leads to the circuit being measured.

-  Read the displayed value. The polarity of red test lead connection  

  will be indicated when making a DCV measurement.

-  When only the figure “1” is displayed, it indicates overrange 

  situation and the higher range has to be selected 

 To avoid electrical shock and/or damage to the instrument, do 

not attempt to take any voltage measurement that might exceeds 

600V dc or ac rms. Do not apply more than 600V dec or ac rms 
between the common terminal and the earth ground.

Current measurement

-  Set the rotary switch to the desired A 

 or A~ range.

-  Connect the black test lead to the COM terminal and the red 

  test leads to the mA terminal for a maximum of 250,A. For a  

  maximum of 10A, move the red test lead to the 10A terminal.

-  Connect test leads in series with the load in which the current is  

  to be measured.

-  Read the displayed value. The polarity of red test lead connection  

Отзывы: