Care of outer surface
Operating Instructions XIOS
Plus
Sensors
7.3
Sterilizing
båÖäáëÜ
The following should be observed regarding XIOS sensors and their
connecting leads:
● Sensors must not be thermally disinfected!
● Sensors must not be immersed in disinfectant solutions!
● The sensors must not be disinfected or sterilized with radiation!
Preliminary Cleaning
Visible contamination
✔
Sensor or cable visibly contaminated (e.g. with blood or saliva).
1. Clean the sensor and cable with a cloth or a paper towel moistened
in soapsuds.
2. Then dry the sensor and cable with a fiber-free cloth or paper towel.
Disinfection
1. Wipe the sensor and cable completely and thoroughly at least twice
with one of the disinfectants recommended in the section "Care and
cleaning agents.
2. Remove any chemical residues by wiping off the sensor with a moist
cloth or gauze.
3. Dry the sensor and the cable with a fiber-free cloth if necessary.
4. Store the sensor for the next treatment in a clean place.
7.3
Sterilizing
Sterilizing
General
NOTICE
Never spray with disinfectants or cleaning agents!
NOTICE
Prior to wipe disinfection or preliminary cleansing, pull the XIOS
Plus
sensor out of the USB module.
NOTICE
Do not let the cable connector of the sensor get wet!
NOTICE
Risk of destruction of the XIOS
Plus
sensors
The XIOS
Plus
sensors are not sterilizable!