63 74 198 D3545
D3545
.
201.06.03
.
02 05.2014
41
Sirona Dental Systems GmbH
Operation
Operating Instructions SIROLaser Advance
Main home screen
båÖäáëÜ
When you activate the finger or wireless foot control, the laser begins
emitting. At the same time, two yellow LEDs at the upper right and left end
of the SIROLaser Advance control unit light up as well as the laser active
bar on the touch screen and the audible alarm sounds. When you release
the finger or wireless foot control to interrupt treatment, the laser is
deactivated, but remains ready for operation.
The following is a typical example of a treatment submenu.
1. Selected program
in this example Frenectomy
2. Laser power
in our example, 3.0 W power. By pushing on the power touch field you will
be transferred to another screen where you will be able to adjust the
emitted power between 0.5 W and 7 W in 0.1 W increments either by
typing in the numbers or by moving higher or lower by using 'plus' or
'minus'.
3. Time
in our example, continuous. The laser will be activated as long as the
finger or wireless foot control is pressed. By pushing on the time touch
field you will be transferred to another screen where you will be able to
adjust the time between continuous or 1 to 9999 seconds either by typing
in the numbers or by moving higher or lower by using 'plus' or 'minus'.
WARNING
All persons present in the room (operator, assistants and patient) must
wear the laser protective goggles as soon as it is advised to wear laser
protective goggles and whenever the green LEDs are lit.
Any actuation of the finger or wireless foot control activates the laser
unit.
Wrong settings may result in severe damage of the patient's soft or hard
tissue or may result in no treatment efficacy. So any user is supposed to
have sufficient knowledge and training in laser therapy.
The treatment room must be protected by suitable measures (in
compliance with IEC 60825-1), e.g. by closing the doors.
NOTICE
Before starting a laser treatment in battery operation please reconfirm
the battery status.
WARNING
The preset power levels are considered to be safe for patients.
Increasing the power levels entails the risk of overheating the patient's
soft or hard tissue. Setting the power to excessively low levels may
result in reduced treatment efficacy.
Содержание SIROLaser Advance
Страница 2: ......
Страница 3: ...MRKOMNQ pfoli ëÉê Çî åÅÉ kÉï ë çÑW båÖäáëÜ léÉê íáåÖ fåëíêìÅíáçåëI å Çá å îÉêëáçå ...
Страница 4: ......
Страница 83: ......
Страница 84: ...MRKOMNQ pfoli ëÉê Çî åÅÉ kçìîÉ ì ÇÉéìáëW cê å áë kçíáÅÉ ÇDìíáäáë íáçåI sÉêëáçå Å å ÇáÉååÉ Notice d utilisation ...
Страница 85: ......
Страница 167: ......
Страница 169: ......