Operation
Sirona Dental Systems GmbH
Main home screen
Operating Instructions SIROLaser Advance
5.5.5.2
Instructions
There occured an error
because the user wants to
set the average power
below 0.05 W in the
chopped mode. Due to
physical limits, this is not
possible. Confirmation via
'OK'.
There occured an error
because the user wants to
set the average power
below 0.3 W in the peak-
pulse mode. Due to
physical limits, this is not
possible. Confirmation via
'OK'.
Internal calibration check:
1 Watt level is being
checked when finger switch
is pressed for three
seconds. Confirmation via
'OK'.
For informations of the cali-
bration check, see chapter
Internal calibration check:
3 Watts level is being
checked when finger switch
is pressed for three
seconds. Confirmation via
'OK'.
Internal calibration check:
5 Watts level is being
checked when finger switch
is pressed for three
seconds. Confirmation via
'OK'.
Internal calibration check:
7 Watts level is being
checked when finger switch
is pressed for three
seconds. Confirmation via
'OK'.