59 38 399 D3352
D3352.076.01.13.02
07.2008
4 – 31
Tab 4
4.3 Adjusting the panoramic X-ray unit
4.
3
Overwriting the default values for Sx
and Sy
z
Overwrite the default values for
Sx
and
Sy
with the measured values in
the text boxes of the
PAN
-
DIAPHRAGM
submenu.
NOTE
i
For information on the direction of displacement (input of +/- sign in the menu)
see page 4-7. Use points as decimal separators!
z
Proceed with step 5 of the adjustment procedure.