6 Program Values – Tele-Radiography

Operating Instructions ORTHOPHOS 3 Ceph

60 82 387 D 3285
D 3285.201.05.04.02

17

båÖäáëÜ

Program Values – Tele-Radiography

a.p. and p.a. Exposures

Encrease kV/mA/s-values for a.p. and p.a. exposures by 
2 steps.

Radiography of Carpus

Recommended value for radiography of carpus: 60kV/
10mA/0,16s.

Film and Processing Notes

The programmed values apply to maximum-sensitivity 
films as Kodak T-Mat G and Agfa Ortholux films in 
combination with the Kodak Lanex Regular intensifying 
screen.

Process according to the manufacturer's instructions of 
the developing equipment and instructions on the 
chemicals.

6.1

Program Values – Tele-Radiography world-wide 
(except USA and Asia)

Intensifying Screen Kodak Lanex Regular with film 
Kodak T-Mat G or Agfa Ortholux.

The differences of density resultant from film and 
development tolerances may be compensated by 
changing the exposure data in one step up or down. 

Table 2:

K

Index

 30

Factory-programmed values 
with a film density of 03

Freely programmed values or values with 
other film density: .... – please enter here –

kV

66

70

74

78

kV

mA

10

10

10

10

mA

s

0.4

0.5

0.64

0.8

s

Table 3:

Possible kV/mA combinations

60

62

64

66

68

70

72

74

76

78

80

kV

 

10

10

10

10

10

10

10

10

10

10

10

mA

 

Table 4:

Possible exposure times in sec.

0.10

0.12

0.16

0.20

0.25

0.32

0.40

0.50

0.64

0.80

1.00

1.20

s

 

1.60

2.00

2.50

3.20

4.00

Содержание orthophos 3 ceph

Страница 1: ...b loqelmelp P l f q J b...

Страница 2: ...ease take the ORTHOPHOS 3 Operating Instructions into account also as well as the Warnings and Safety Notes included therein Ihr ORTHOPHOS Team Maintenance To ensure the safety of the patient the oper...

Страница 3: ...r 8 3 Accessories 9 3 1 Accessories for Cephalometer 9 3 2 Adjusting Film 9 4 Operation 10 4 1 Preparing the Tele Exposure 10 4 2 Positioning the Patient for Tele Exposures 12 4 3 Releasing the Tele E...

Страница 4: ......

Страница 5: ...th the regulations stated in the Directive 93 42 EEC of June 14 1993 concerning medical products Exposure times tele exposure 0 1 4 00s Automatic Exposure Blockage The duration of the exposure blockag...

Страница 6: ...witch 2 Patient positioning mirror 3 Head holder with adjusting knob for temple supports 4 Cassette holder with carriage 5 Diaphragm 6 Diaphragm wheel with locking button 7 Height adjustment buttons A...

Страница 7: ...splay for exposure program exposure time with buttons for exposure programs Exposure time for tele radiography Digital display for kV mA paired values with keys for overriding kV values Patient symbol...

Страница 8: ...25 Projection scale 26 Nose support position scale horizontal 27 Ear olives with holders 28 Film cassette for tele radiography Reproducibility of Tele Exposures After positioning the patient note the...

Страница 9: ...W may be disinfected with petroleum ether We recommend that you attach hygienic protective covers before each exposure 3 2 Adjusting Film The supplied adjusting film is needed for the maintenance and...

Страница 10: ...mmetrical S for a p p a and carpus exposures Adjust diaphragm by pressing locking button and turning diaphragm wheel up to stop The diaphragm number appears in the upper right corner of the diaphragm...

Страница 11: ...top and push out to side The cephalometer can be rotated The rotational apparatus locks in the 0 45 turning to the right and 90 position 0 position is for lateral exposures For adjustment the swivel l...

Страница 12: ...or sit under cephalometer Press buttons or on unit to adjust cephalometer until both ear olives are at same height as external auditory canals Introduce ear olives into external auditory canals Hold...

Страница 13: ...rding set downwards Selecting the tele exposure data Briefly press one of the four program keys The LED above the key illuminates The corresponding programmed kV mA values and exposure time s are disp...

Страница 14: ...5 04 02 Slide film cassette into cassette holder up to lateral cassette stop locks facing cassette holder not pati ent Front cassette stop A asymmetric for lateral projections Back cassette stop S sym...

Страница 15: ...ED goes out If no further exposures are to be made switch unit off Interrupting an Exposure If the exposure key is released prematurely exposure is interrupted The kV mA displays and the LED above key...

Страница 16: ...on page 17 1 Adjust diaphragm 3 or 4 for ceph function The C dis play above s on the Multitimer illuminates 1 Set desired kV and s values with keys 2 Press memory programming key The LED above the pr...

Страница 17: ...e developing equipment and instructions on the chemicals 6 1 Program Values Tele Radiography world wide except USA and Asia Intensifying Screen Kodak Lanex Regular with film Kodak T Mat G or Agfa Orth...

Страница 18: ...m film and development tolerances may be compensated by changing the exposure data in one step up or down Table 5 Index 2A Factory programmed values with a film density of 03 Freely programmed values...

Страница 19: ...om film and development tolerances may be compensated by changing the exposure data in one step up or down Table 8 Index 1A Factory programmed values with a film density of 03 Freely programmed values...

Страница 20: ...t K p a p d e OMMN p W m d a POURKOMNKMRKMQKMO MPKOMMT KJk KW NMU ROP f p a p d e rp W W c PN SQSOR _ d K K p a p ii QUPR p a I p NMM I k OUOTP rp p PORM o a J r R j I l iRi RvS l k SM UO PUT a POUR...

Отзывы: