Button Functions
Button
Function
To turn the meter on and enter the blood
glucose test result review mode.
To change sample type of test result
review
To turn the meter off.
To delete test results.
To confirm date/time settings.
To adjust settings for date-time/ scroll
through test results.
To decrease the number of memory when
review test result.
Short press ▼.
To increase the number of memory when
review test result
Press and release ▼.
To adjust settings for date-time/ scroll
through test results.
Press and release ▲.
Short press ▲.
▲
▼
Press and hold for 3 seconds.
Action
Short press
Press and hold for 3 seconds.
In test result review mode, press and ▼
together to delete all kinds of test result .
Short press
Setting Up Your Safe AQ UG Blood Glucose and Uric Acid meter
Setup steps for date/time
The Safe AQ UG blood glucose and uric acid meter is preset with the date and time. You may
need to adjust it to your local time zone. Check the date and time each time the batteries are
replaced. Reset the date and time if they are not correct.
STEP 1
: Insert the Batteries
Open the Battery Cover on the back side of the Safe AQ UG blood glucose and uric acid
meter. Insert two AAA alkaline batteries as indicated by the “+” and “-” symbols.
STEP 2
: Set Correcting Code of Uric Acid (use code chip)
Code chip matching with test strips shall be used when you use a new vial of uric acid test
strip. Find code chip in test strip package, check if the code number on code chip is in
accordance with code number on test strip package. If not, you will get inaccurate test
results, please call Customer Service at +86-731-8986-731-89935582. If there is
already a code chip in Safe AQ UG meter, please take it out and insert a new code chip.
Insert the code chip to code chip port, the meter will make a short beep sound and perform
self-inspection. The meter will display code number and meter will be off after a short beep
sound.
If there is an error, the meter will display E-4. Please pull out the code chip and insert again. If
still failed, please call Customer Service at +86-731-8986-731-89935582.
Every code chip only need to be inserted once, the meter will store the code number until
insert new code chip.