Power Saving
If there is no blood applied to the test strip within 3 minutes, the meter will automatically
turn off.
After test, the test results will be displayed on the screen with test strip in strip port, If there
no operation on the meter within 3 minutes, the meter will automatically turn off.
If there is no other operation on the meter within 1 minute, such as enter into test results
review mode, the meter will automatically turn off.
Specifications
Performance Characteristics
Accuracy: 95%test result of glucose meet the requirement below:
95% test result of uric acid meet the requirements below:
Blood volume for blood glucose
About 0.6μL
Blood volume for uric acid
About 3μL
Sample type
Capillary whole blood, Venous whole blood
Calibration
Plasma equivalent
Measuring time for blood glucose
5±1s
Measuring time for uric acid
25±1s
Meter storage/transportation conditions
-20℃~55℃
<5.5 mmol/L(100 mg/dL)
Concentration Range
Bias%
Within ± 0.83 mmol/L(± 15mg/dL)
Within ± 15%
≥5.5 mmol/L(100mg/dL)
Precision: test result of glucose meet the requirement below:
<5.5 mmol/L(100 mg/dL)
Concentration Range
Requirement
SD<0.34mmol/L(6 mg/dL)
CV<6.0%
≥5.5 mmol/L(100mg/dL)
≤297 μmol/L(5 mg/dL)
Concentration Range
Bias%
≤ ± 59.4 μmol/L(1 mg/dL)
≤ ± 20%
>297 μmol/L(5 mg/dL)
Test result of uric acid meet the requirements below:
Details please refer to blood glucose and uric acid package insert.
≤297 μmol/L(5 mg/dL)
Concentration Range
Requirement
SD < 22.2 μmol/L(0.37mg/dL)
CV < 7.5%
>297 μmol/L(5 mg/dL)