Low battery power. 

Meter is out of the required 
testing temperature range. 

Replace the batteries immediately.

Place the meter and test strip at the 

operating temperature range.

Repeat the test when the meter and 

test strips have reached a 

temperature within the operating 

range.

Display

What it means

What to do

Used  blood glucose test strip

Expired test strip or damped 

test strip 

For uric acid testing

Used test strip

Use a new test strip to test again.

Test method error: insert strip 
into meter after apply sample 
to strip.

Use a new test strip to test again.
Use a new test strip to test again.

Test strip exposed to air more 

than 3 minutes.

Use a new test strip to test again.

Insert test strip into meter before 
apply sample to strip.

Display

What it means

What to do

For blood glucose testing

Test method error: insert strip 
into meter after apply sample 
to strip.

Insert test strip into meter before 
apply sample to strip.

Your test result is higher than 
20.0 mg/dL or 1188μmol/L.

Use a new test strip to test again, if 

meter still display “HI”, call healthcare 

professional immediately.

Содержание Safe AQ UG

Страница 1: ...nostic use only Safe AQ UG Blood Glucose and Uric Acid Meter Blood Glucose and Uric Acid Monitoring System Benutzerhandbuch Nur zur In vitro Diagnostik Safe AQ UG Blutzucker und Harns ure Blutzucker u...

Страница 2: ...monitoring system This user manual contains important information about the meter and how it works Please read this user manual carefully before using your new meter or the protection provided by the...

Страница 3: ...our Safe AQ UG Blood Glucose and Uric Acid meter Control Solution Testing Blood Glucose and Uric Acid Testing Viewing test result Memory Information for Healthcare Professional Use Comparing Meter wit...

Страница 4: ...and uric acid meter Safe AQ UG blood glucose test strip Safe AQ UG uric acid test strip blood glucose control solution uric acid control solution Unit type Safe AQ UG Blood Glucose and Uric Acid meter...

Страница 5: ...e choking hazards DO NOT drink the control solution Remove the batteries if the meter is not likely to be used for some time DO NOT CHANGE YOUR TREATMENT BASED ON A SINGLE RESULT THAT DOES NOT MATCH H...

Страница 6: ...low Plasma Blood Glucose Result 100 mg dL 5 55 mmol L Fasting Plasma Glucose Male 202 mol L 416 mol L 3 4 mg dL 7 0 mg dL 142 mol L 339 mol L 2 4 mg dL 5 7 mg dL Female 1 If blood glucose test results...

Страница 7: ...he blood glucose test strips here Main Button Power Set Turns the Meter on o Sets selected parameter Enter test results memory mode Up arrow Increase data LCD Display The display shows results message...

Страница 8: ...with the date and time You may need to adjust it to your local time zone Check the date and time each time the batteries are replaced Reset the date and time if they are not correct STEP 1 Insert the...

Страница 9: ...e meter will turn on The meter enters into the Date Time mode 2 It displays date and time the segment of year flashes Step 1 With the meter turned o press and hold for 3 seconds to start Time and Date...

Страница 10: ...onth Step 1 Press and release or to decrease or increase the month Step 2 Short press and release the main button to con rm the month STEP 6 Setting the Day The day appears flashing at the left side o...

Страница 11: ...crease or increase the hour Step 2 Short press and release to con rm the hour STEP 8 Setting the Minutes The minutes appear flashing at the lower left side of meter display 1 Press and release or to d...

Страница 12: ...tion Testing Why Perform a Control Solution Test Ensures that your meter and test strips are working properly Allows you to practice testing without using blood When Should You Perform a Control Solut...

Страница 13: ...g blood drop symbol and control bottle symbol can you apply control solution If there is no blood drop symbol or control bottle symbol restart the meter again or change the test mode STEP 3 Apply the...

Страница 14: ...test Temperature is too high or too low Wrong correcting code Equilibrate meter test strip and control solution in room temperature for 30 minutes and retest Set correcting code which accordance with...

Страница 15: ...ge skin 5 deepest depth for thick or calloused skin The Sinocare lancing device is prepared and ready to lance your nger for a blood sample The lancet is for single use only To prevent the risk of blo...

Страница 16: ...will not turn on The meter can test blood glucose and uric acid together STEP 10 Take Blood Sample Make sure the lancing device is pressed rmly to your nger Press the button to lance Attention If you...

Страница 17: ...ric acid testing After a 25 second count down the meter displays the test result in mg dL or mol L STEP 13 Eject the strip Used test strip and lancet may be biohazardous Please discard them carefully...

Страница 18: ...ter than 33 3mmol L 600 mg dL the meter will display HI Retest immediately with a new test strip If your reading is still HI contact healthcare professional immediately If blood glucose test results l...

Страница 19: ...sult is dropped and the newest is added Press the or button to view all the blood glucose results in memory When you have viewed all test results meter will display End press and hold the for 3 second...

Страница 20: ...professional be allowed to collect venous blood samples Any patient who has infectious diseases must use his her own meter If the same meter be used on multiple patients there is risk of infection bet...

Страница 21: ...oughly again after handing meter 5 Verify the meter is working properly by performing control solution test If your meter can t work properly please contact customer service at 86 731 89935581 86 731...

Страница 22: ...cose About 0 6 L Blood volume for uric acid About 3 L Sample type Capillary whole blood Venous whole blood Calibration Plasma equivalent Measuring time for blood glucose 5 1s Measuring time for uric a...

Страница 23: ...ay cause erroneous results Electromagnetic Compatibility EMC The meter complies with the electromagnetic requirements IEC 61326 2 6 2012 speci ed in ISO 15197 2013 Electromagnetic emissions are low an...

Страница 24: ...sional immediately Use a new test strip to test again if meter still display LO call healthcare professional immediately Display What it means What to do Your blood glucose test result is below the me...

Страница 25: ...p to test again Test method error insert strip into meter after apply sample to strip Use a new test strip to test again Use a new test strip to test again Test strip exposed to air more than 3 minute...

Страница 26: ...right port when insert the test strips The blood glucose test strip insert into port marked with G The uric acid test strip insert into port marked with U Test strip Inserts into wrong ports Poor con...

Страница 27: ...new test strip The test strip is exposed in moisture environment for too long Display What it means What to do without sample applied to test starts and meter counts down as soon as the test strip is...

Страница 28: ...ard carefully Use 75 alcohol to clean your meter and send your meter with warranty card to your local dealer References 1 American Diabetes Association Standards of medical care in diabetes 2016 2 Lar...

Страница 29: ...intenance 40 Change battery 41 Troubleshooting 42 Symbol 49 Warranty 50 Thank you for choosing Safe AQ UG blood glucose meter a free replacement under normal operation is provided Please do complete t...

Страница 30: ...53...

Страница 31: ...Tech Zone Changsha Hunan Province 410205 People s Republic of China Tel 86 731 89935581 89935582 Fax 86 731 89825189 Email info sinocare com Website www sinocare com Ei estra e 80 20537 Hamburg German...

Отзывы: