Slide Switch:

This slide switch is used to select the response time of the

meter circuit as defined in Table 1-1.

3.

BATT  TEST

A momentary push-button switch used to check the condi-

tion of the battery.

4.

Microphone

A condenser type microphone that converts sound en-

ergy vibrations into electrical signals.

5.

Output Jacks

Decibel Output
Jack:

A logarithmic output of 1.50 V DC is equivalent to meter
reading of +10 dB looking into 100  K

 load. The recom-

mended plug of this jack is a Switchcraft #850.

RMS Output
Jack:

A weighted or non-weighted output of 1.00 V RMS is equiva-
lent to a meter reading of +10 dB looking into a 100 K

load. The recommended plug for this jack is a Switchcraft
#850.

External Filter
Jack:

A non-weighted (flat response) output of 120 millivolts RMS
is equivalent to a meter reading of +10 dB looking into 100
K

 load. The recommended plug for this jack is a Switchcraft

#750.

6.

Meter:

A 0-100 μA DC full scale, taut-band movement contains a

dB scale with end markings of –10 dB and +10 dB and 1
dB divisions starting from –5 dB. The upper end of the
scale includes a BATT OK section.

FAST

SLOW

(OSHA)

SOUND LEVEL METER

884-2

TYPE S2A

BATT
TEST

O

S

H

A

130

120

100

110

90

80

70

60

50

OF

F

5

3

8

6

7

1

2

4

Figure 4-1. Operational Items
and Features

www.

.com

[email protected]

1.800.561.8187

Содержание 884-2

Страница 1: ...Simpson Model 884 2 Sound Level Meter OPERATOR S MANUAL www com information itm com 1 800 561 8187...

Страница 2: ...nt and other articles manufactured by it to be free from defects in material and workmanship under normal use and service its obligation under this warranty being limited to making good at its factory...

Страница 3: ...LICATION 11 3 1 Decibel Scale 11 3 2 Measurement And Exposure Standards 11 3 3 Sound Fields 12 3 3 1 Free Field 12 3 3 2 Reverberant Field 12 3 3 3 Semi Reverberant Field 13 3 4 Microphone Orientation...

Страница 4: ...r switch which references the 0 dB point of the indicating instrument to the range selected For example when the range switch is set to 70 the zero 0 mark on the black arc is equal to 70 dB Fast or sl...

Страница 5: ...sure levels of either 94 dB on the 90 dB range or 114 dB on the 110 dB range Screwdriver adjustable from the side of the case 9 Output Connections This Instrument is provided with three outputs Exter...

Страница 6: ...Windscreen 00197 Calibrator 890 2 12890 Package Case 10 866065 Adapter Microphone to Tripod 00184 898 Octave Band Filter Set 02231 2 INSTALLATION This section contains information and instructions for...

Страница 7: ...ached to any commercial photography type tripod having a standard 1 4 20 mounting screw The mounting screw is aligned with the threaded insert in the case and finger tightened A camera type carrying s...

Страница 8: ...st that can be tolerated without pain Because of the large ratio of value involved a linear meter scale marked in pressure units becomes impractical A scale based on logarithms called a decibel dB sca...

Страница 9: ...are established by the Oc cupational Safety and Health Act of 1970 OSHA the Walsh Healy Act Federal Register Volume 34 No 96 and paragraph 1910 95 of the Occupational Safety and Health Standards Feder...

Страница 10: ...sence When sound is coming mainly from one direction the reading might be affected by the relative positions of the microphone and operator For example if the operator and microphone are in line and f...

Страница 11: ...tergent Thoroughly rinse and dry before placing over microphone 3 7 Meter Fluctuations If the meter indication fluctuates more than 3 dB set the meter response switch of the Instrument to the SLOW pos...

Страница 12: ...9 Dual Sound Sources When separate measurements are made on two sound sources the combined effect cannot be predicated by direct addition For example If two machines are emitting equal noise levels of...

Страница 13: ...und Level Meter periodically especially if the Instrument has been subjected to severe me chanical shock or environmental extremes 4 2 Operational Items And Features All items and features used to ope...

Страница 14: ...f 1 00 V RMS is equiva lent to a meter reading of 10 dB looking into a 100 K load The recommended plug for this jack is a Switchcraft 850 External Filter Jack A non weighted flat response output of 12...

Страница 15: ...her applications select the desired range weighting and response If the meter fluctuations are in excess of 3 dB set the response to SLOW position NOTE With the range switch set to any of the other ra...

Отзывы: