Slide Switch:
This slide switch is used to select the response time of the
meter circuit as defined in Table 1-1.
3.
BATT TEST
A momentary push-button switch used to check the condi-
tion of the battery.
4.
Microphone
A condenser type microphone that converts sound en-
ergy vibrations into electrical signals.
5.
Output Jacks
Decibel Output
Jack:
A logarithmic output of 1.50 V DC is equivalent to meter
reading of +10 dB looking into 100 K
⍀
load. The recom-
mended plug of this jack is a Switchcraft #850.
RMS Output
Jack:
A weighted or non-weighted output of 1.00 V RMS is equiva-
lent to a meter reading of +10 dB looking into a 100 K
⍀
load. The recommended plug for this jack is a Switchcraft
#850.
External Filter
Jack:
A non-weighted (flat response) output of 120 millivolts RMS
is equivalent to a meter reading of +10 dB looking into 100
K
⍀
load. The recommended plug for this jack is a Switchcraft
#750.
6.
Meter:
A 0-100 μA DC full scale, taut-band movement contains a
dB scale with end markings of –10 dB and +10 dB and 1
dB divisions starting from –5 dB. The upper end of the
scale includes a BATT OK section.