10 Push-button (8455 V3)

 This push-button is used for resistance measurements only. While the button is
depressed, the resistance range becomes 1/10 of the dial markings and the
indication is then divided by 10. For example, if the value indicated is less than 10
K

V

, the scale divisions are very small. By depressing the R

4

10 button, 0-10 K

V

 is

spread across 1/2 of the scale and a more accurate reading can be obtained. If, in
a ballistic test, the meter pointer kicks off-scale (past 100 POINTS), holding the
R

4

10 button will reduce the “kick” to a readable value.

10.5

Reverse Switch

This switch is primarily for convenience when making ballistic tests. Operating
the switch back and forth transposes the test leads to produce the capacitive
charging “kick” when connected to an open line conductor. It can also be used to
reverse the polarity of the test leads for voltage measurements.

NOTE:

 The Instrument can be operated in any convenient position; however,

maximum reading accuracy is obtained with panel in horizontal position as the
Instrument was calibrated in this position.

11.

TEST FOR GROUND

a.

Open conductors toward
Central Office.

b.

Set the Instrument to mea-
sure resistance (ON/OFF
switch to ON; “Press for
Ohms” depressed, 8455A).

c.

Short test clips together and
adjust BATTERY ADJUST-
MENT control for meter
reading of full scale (100
POINTS).

d.

Connect one test clip to ground*.

*GROUND: A connection to metallic cable sheath or wrap, to cable suspension
strand, to ground rod, or to grounded (T) conductor of an adjacent pair.

e.

Connect the other clip to conductor giving a meter deflection.

f.

Note the exact value of resistance indicated. For resistance less than 10 k

V

,

hold R

4

10 button depressed and divide indicated value by 10 (8455 V3).

g.

To calculate distance to fault, refer to Section 17.

12.

TEST FOR SHORT CIRCUIT

a.

Open conductors toward Central Office.

b.

Set the Instrument to measure resistance (ON/OFF switch to ON; “Press for
Ohms” depressed, 8455A).

c.

Short test clips together and adjust BATTERY ADJUSTMENT control for meter
reading of full scale (100 POINTS).

d.

Check both conductors for grounds.

e.

Connect test clips across line pair.

CONDUCTORS OPENED FOR TEST

R

T

FAULT

R

T

TO

STATION

TO

CENTRAL

OFFICE

ON

8455

GROUND 

*

Figure11- 1. Test for Ground

www.

.com

[email protected]

1.800.561.8187

Содержание 8455 V3

Страница 1: ...Simpson 8455 V3 And 8455A Line Loop Testers OPERATOR S MANUAL Model 8455 V3 Model 8455A www com information itm com 1 800 561 8187...

Страница 2: ...nt and other articles manufactured by it to be free from defects in material and workmanship under normal use and service its obligation under this warranty being limited to making good at its factory...

Страница 3: ...R4 10 Push button 8455 V3 10 10 5 Reverse Switch 10 11 TEST FOR GROUND 10 12 TEST FOR SHORT CIRCUIT 10 13 TEST FOR CROSSED CONDUCTORS 11 14 TEST FOR OPEN 11 15 TEST FOR SPLIT 12 15 1 Identify Split Co...

Страница 4: ...from the 8455 V3 in that it substitutes a 0 10K range for the 0 200K range of the 8455 V3 The 2M range remains for points measurements but does not have a separate meter dial scale The 8455 V3 differs...

Страница 5: ...or supplier immediately Check that all items and accessories are included with the instrument as listed in Table 2 1 5 ELECTRICAL CHECKOUT After determining that the 8455 is undamaged check the Instr...

Страница 6: ...d it carefully before requesting any warranty repairs NOTE For all assistance contact the nearest Simpson Authorized Service Cen ter 7 SHIPPING Pack the Instrument carefully and ship it prepaid and in...

Страница 7: ...ly de energized 8 Case Assembly Screws 4 Screws to be removed to access battery 9 SAFETY PRECAUTIONS The following safety symbols may be found in this manual and or on the Instru ment They call attent...

Страница 8: ...n metallic dust can be explosive 7 Do not use this Instrument to make electrical measurements on blasting circuits or blasting caps 10 OPERATION 10 1 Mechanical Zero Adjustment For accuracy of all mea...

Страница 9: ...strument was calibrated in this position 11 TEST FOR GROUND a Open conductors toward Central Office b Set the Instrument to mea sure resistance ON OFF switch to ON Press for Ohms depressed 8455A c Sho...

Страница 10: ...to mea sure resistance ON OFF switch to ON c Connect one test clip to ground d Connect the other test clip to the good conductor not open e Operate the REVERSE switch back and forth slowly allow time...

Страница 11: ...line POINTS Distance To Open 100 100 of line length 90 80 of line length 80 60 of line length 60 45 of line length 40 30 of line length 20 15 of line length 10 10 of line length 5 5 of line length 15...

Страница 12: ...sent on a line conductor with respect to ground or by connecting the clips to both conductors any voltage present across the line 100 POINTS 100 volts NOTE The Instrument will not indicate the presenc...

Страница 13: ...scribes how to maintain the Instrument in good operating condi tion This Instrument is carefully designed and constructed with high quality com ponents By providing reasonable care and following instr...

Страница 14: ...condition can be checked on the 8455 V3 by setting the ON OFF switch to OFF REVERSE switch to normal and touching red test lead clip to battery terminal A meter indication of at least 35 POINTS 35 V...

Страница 15: ...r transport should be checked frequently for specified accuracies If such facilities are not available the Instrument may be sent to the factory Service Department for an overall checkout repair if ne...

Отзывы: