background image

9

1.

The 8455 should only be used by personnel who are qualified to recognize
shock hazards and are trained in the safety precautions required to avoid
possible injury.

2.

Do not connect any terminal of this Instrument to a circuit point at which a
voltage exceeding 100 volts DC may exist with respect to earth ground.
(Refer to Table 8-1, item 4).

3.

This Instrument is intended only for indoor use or in sheltered outdoor loca-
tions. To prevent fire or shock hazards, do not expose Instrument to rain or
moisture.

4.

Before using this Instrument, check both the Instrument and the test leads for
missing, damaged, deteriorated or faulty insulation. Do not use the instru-
ment until it has been properly repaired.

5.

DC voltages under 60V are not considered to be a shock hazard; however,
the meter is capable of measurements up to 100V. If the Instrument is to be
used to measure voltages exceeding 60V DC, precautions should be taken
to avoid bodily contact with the energized circuit.

6.

Do not make electrical measurements where the air may contain explosive
concentrations of gas or dust such as in mines, grain elevators, gasoline
stations or in the presence of charging batteries, until determined to be safe
by qualified personnel. Note that even metallic dust can be explosive.

7.

Do not use this Instrument to make electrical measurements on blasting
circuits or blasting caps.

10.

OPERATION

10.1

Mechanical Zero Adjustment

For accuracy of all measurements, the meter pointer must rest directly over the 0
POINTS mark at the left end of the dial scale  when the ON/OFF switch is set to the
OFF position. If the pointer is not in that position, rotate the mechanical zero adjust
screw (Figure 8-1, item 7) in either direction until it does.

10.2

Battery Adjustment (Resistance Ranges)

For accuracy of resistance measurements, the meter pointer must rest directly
over the 100 POINTS mark at the right end of the dial scale when the ON/OFF
switch is set to the ON position and the test clips are shorted together. If the pointer
is not in that position, adjust the BATTERY ADJUSTMENT control (Figure 8-1,
item 5).

10.3

Storage and/or Transport

Keep the ON/OFF switch in the OFF position at all times except when making
measurements. This will prevent battery discharge in the event that the test clips
are accidentally shorted together during storage or transport.

NOTE:

 Before making a resistance or ballistics test, connect the Instrument to the

line with the ON/OFF switch in the OFF position (voltage measurement) to detect
the presence of unexpected DC voltages. The presence of DC voltages will cause
inaccurate test readings and may damage the meter.

www.

.com

[email protected]

1.800.561.8187

Содержание 8455 V3

Страница 1: ...Simpson 8455 V3 And 8455A Line Loop Testers OPERATOR S MANUAL Model 8455 V3 Model 8455A www com information itm com 1 800 561 8187...

Страница 2: ...nt and other articles manufactured by it to be free from defects in material and workmanship under normal use and service its obligation under this warranty being limited to making good at its factory...

Страница 3: ...R4 10 Push button 8455 V3 10 10 5 Reverse Switch 10 11 TEST FOR GROUND 10 12 TEST FOR SHORT CIRCUIT 10 13 TEST FOR CROSSED CONDUCTORS 11 14 TEST FOR OPEN 11 15 TEST FOR SPLIT 12 15 1 Identify Split Co...

Страница 4: ...from the 8455 V3 in that it substitutes a 0 10K range for the 0 200K range of the 8455 V3 The 2M range remains for points measurements but does not have a separate meter dial scale The 8455 V3 differs...

Страница 5: ...or supplier immediately Check that all items and accessories are included with the instrument as listed in Table 2 1 5 ELECTRICAL CHECKOUT After determining that the 8455 is undamaged check the Instr...

Страница 6: ...d it carefully before requesting any warranty repairs NOTE For all assistance contact the nearest Simpson Authorized Service Cen ter 7 SHIPPING Pack the Instrument carefully and ship it prepaid and in...

Страница 7: ...ly de energized 8 Case Assembly Screws 4 Screws to be removed to access battery 9 SAFETY PRECAUTIONS The following safety symbols may be found in this manual and or on the Instru ment They call attent...

Страница 8: ...n metallic dust can be explosive 7 Do not use this Instrument to make electrical measurements on blasting circuits or blasting caps 10 OPERATION 10 1 Mechanical Zero Adjustment For accuracy of all mea...

Страница 9: ...strument was calibrated in this position 11 TEST FOR GROUND a Open conductors toward Central Office b Set the Instrument to mea sure resistance ON OFF switch to ON Press for Ohms depressed 8455A c Sho...

Страница 10: ...to mea sure resistance ON OFF switch to ON c Connect one test clip to ground d Connect the other test clip to the good conductor not open e Operate the REVERSE switch back and forth slowly allow time...

Страница 11: ...line POINTS Distance To Open 100 100 of line length 90 80 of line length 80 60 of line length 60 45 of line length 40 30 of line length 20 15 of line length 10 10 of line length 5 5 of line length 15...

Страница 12: ...sent on a line conductor with respect to ground or by connecting the clips to both conductors any voltage present across the line 100 POINTS 100 volts NOTE The Instrument will not indicate the presenc...

Страница 13: ...scribes how to maintain the Instrument in good operating condi tion This Instrument is carefully designed and constructed with high quality com ponents By providing reasonable care and following instr...

Страница 14: ...condition can be checked on the 8455 V3 by setting the ON OFF switch to OFF REVERSE switch to normal and touching red test lead clip to battery terminal A meter indication of at least 35 POINTS 35 V...

Страница 15: ...r transport should be checked frequently for specified accuracies If such facilities are not available the Instrument may be sent to the factory Service Department for an overall checkout repair if ne...

Отзывы: