32 GB/CY
Problem
Possible cause
solution
Pulse oximeter
shows inter-
rupted meas-
urements or
severe jumps in
the measure-
ment.
Circulation problems in the
finger being measured.
Refer to warnings and safety notes.
The finger being measured is
too large or too small.
The fingertip must be the following
size: between 10 and 20 mm wide.
Between 5 and 15 mm thick.
The finger, hand or body are
moving.
Keep the finger, hand and body still
during measurement.
Arrhythmia.
Seek medical attention.
Unable to
transfer the
measurements.
The "HealthForYou" app is not
activated or
Bluetooth
®
is off
in the smartphone settings.
Activate
Bluetooth
®
on your
smartphone and start the app.
The batteries in the pulse
oximeter are low or drained.
Replace the batteries.
The device is still on.
Wait for the device to switch off.