420903-
Essential Peel Instructions Manual-rev000
l
7
2.4
Unit Preparation before Microdermabrasion
Before each session, you MUST check the following:
§
The clean crystal bottle is tightly in place.
§
The "used crystal" container has been emptied or replaced.
§
Before putting container in place, remove any crystal residue from the ridges with a dry tissue and make sure
that the container is properly closed.
WARNING: For sanitary reasons and for the safety of your unit, corundum crystals must never be reused.
Starting procedure session
1.
Turn on the main power switch on the back of the unit (#1).
The LED indicator on front panel will light up.
2.
Wait at least 10 seconds for unit to warm up.
3.
Press the “START / STOP” key (#2 on the front panel)
.
4.
Adjust the abrasion/vacuum power with the intensity key (#3 on the front panel)
.
5.
The intensity will appear on your screen (#13).
6.
Start exfoliation procedure.
2.5
Filters Replacement
The Essential Peel unit has 2 filters which
require
periodic
replacement.
Skin debris vacuumed into the “used crystal”
container
reduces
efficiency. When suction diminishes, it is time to change the filter*.
NOTE:
The filters will require less maintenance if the client’s skin is perfectly cleaned and dried prior to each session. We recommend
the use of a water based cleansing
lotion
to avoid clogging problems The skin must be completely free of any moisture; use a hair dryer
to complete the drying process. Any moisture left on the skin will create hose or filter clogging.
How to change the small “used crystal” filter (ST Code 416597):
Small “used crystal” filter
1.
Rubber ring
2.
Plastic mesh (filter)
Procedure:
§
This small filter is good for approximately 40-50 procedures.
§
When the filter needs to be replaced, “CF” inscription will appear on the display screen (# 13 on the front
panel).
§
Unscrew the used crystal jar.
§
Pull out the used filter and discard it.
§
Holding the new filter by the white plastic base, push it in with a light twist.
§
Press concomitantly on both intensity control arrows (# 3 on the front panel) to reset the “CF” inscription.
§
You can restart a new series of about 40-50 sessions.