2 Configuration 

To change device settings, press 

Menu 

(key 4) for 2 seconds. This will call the configuration menu.  

Pressing key 

Menu

 jumps between the parameters. 

The parameters can be changed with 

5

 (key 2) or 

6

 (key 5).  

Quit

 (key 6) finishes the configuration and returns to standard measuring operation. 

2.1  Unit: Choice Of The Display Unit 

Choose the desired display unit, the referring unit is displayed by means of a functional arrow in 
the display. The selection is permanently stored in the sensor, therefore after reconnecting the 
sensor the unit will instantly reappear. The choice depends on the used sensor. 

2.2  Sea Level Correction For Absolute Pressure Sensors 

The device displays the absolute pressure measured at the sensor. This is not necessarily the same like the 
values given by weather stations! The weather stations‘ values are pressure at sea level. Usually the sensor 
is placed above sea level and therefore, if the value at sea level(zero) is to be measured, the pressure loss 
resulting from the actual level above zero has to be considered! 

ƒ

To correct activate the „Sea-Level-Function“:
Select „on“ in the menu „SL“ with 

5

 (key 2) or 

6

 (key 5).

ƒ

Jump to the next parameter „Alti“ by pressing 

Menu

 (key 4).

Then enter the altitude above sea level of the sensor‘s location in meters and leave the
configuration by pressing

 Quit

 (key 6)

If the sea level correction is active this will be shown by the functional arrow „SL“ in the display, the device 
now displays the absolute pressure at sea level (zero). 

2.3  P.oFF: Auto Power Off Time 

The device will be automatically switched off if no key is pressed/no interface communication 
takes place for the time of the power off time.  
The power off time can be set to values between 1 and 120 min. It can be completely 
deactivated by setting the parameter to ‚P.oFF = oFF“.  

2.4  Adr: Base Address of Interface 

Up to 10 devices of the MH3xxx- handheld-family can be connected to a serial interface at once 
(depending on interface converter, e.g. GRS3105: 5 devices). To get access to each device the 
base addresses of the devices have to be different. For example choose 01 for the first, 11 for 
the second device and so on.  

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

mbar

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

MPa

Pa

bar

mbar

kPa

Tara

AL Logg

SL

PSI

mbar

2.5 OFFS: Adjusting Sensor Zero Displacement 

A zero displacement can be carried out for the measured value: 

value displayed = value measured  -  offset 

Standard setting: 'off' = 0.0°, i.e. no zero displacement will be carried out. Together with the scale correction 
(see below) this factor is mainly used to compensate for sensor deviations. Input is in the display unit. 

2.6 SCAL: Adjusting Sensor Scale 

The scale of the measuring can be influenced by this setting (factor is in %): 

displayed value = measured value * (1+Scal/100) 

Standard setting: 'off' =0.000, i.e. value is not corrected. Together with the zero displacement (see above) 
this factor is mainly used to compensate for sensor deviations. 

SIKA Dr.Siebert & Kühn GmbH & Co. KG ·• Struthweg 7–9 • 34260 Kaufungen • Germany • 

 +49 5605 803-0 • 

 +49 5605 803-54 • 

 [email protected] • www.SIKA.net 

Содержание MH3111

Страница 1: ...r s Manual Handheld Pressure Meter MH3111 as of Version 6 4 for MSD or MXD Pressure Sensors SIKA Dr Siebert K hn GmbH Co KG Struthweg 7 9 34260 Kaufungen Germany 49 5605 803 0 49 5605 803 54 info sika...

Страница 2: ...URE SENSORS 5 2 3 P OFF AUTO POWER OFF TIME 5 2 4 ADR BASE ADDRESS OF INTERFACE 5 2 5 OFFS ADJUSTUNG SENSOR ZERO DISPLACEMENT 5 2 6 SCAL ADJUSTUNG SENSOR SCALE 5 3 MEASURING OF WATER LEVEL DISPLAY UNI...

Страница 3: ...e product as safety or emergency stop device or in any other application where failure of the product could result in personal injury or material damage Failure to comply with these instructions could...

Страница 4: ...max key 2 shows the maximum of the measured value Pressing it again hides them To clear the max memory press key max for 2 seconds Min Memory Pressing min key 5 shows the minimum of the measured valu...

Страница 5: ...s the absolute pressure at sea level zero 2 3 P oFF Auto Power Off Time The device will be automatically switched off if no key is pressed no interface communication takes place for the time of the po...

Страница 6: ...ASYCONTROL Universal multi channel software EASYBUS RS485 or MH3000 operation possible for real time recording and presentation of measuring data of one MH3xxx device in the ACCESS data base format In...

Страница 7: ...System Messages Display Meaning What to do Low battery power device will only continue operation for a short period of time Replace battery Battery empty Replace battery Mains operation without batter...

Страница 8: ...for measuring values and for min max memories hold function etc as well as additional functional arrows Pushbuttons 6 membrane keys Interface Serial interface 3 5mm jack can be connected to RS232 or...

Отзывы: