1 General 

1.1 Safety Requirements 

This device has been designed and tested in accordance with the safety regulations for electronic devices. 
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures 
and special safety advises given in this manual will be adhered to when using the device. 
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected

to any other climatic conditions than those stated under "Specification".

2. Device and sensors have to be handled with care (don’t throw, hit, etc.). Protect plugs and sockets from

soiling.

3. If the device is transported from a cold to a warm environment condensation may cause in a failure of the

function. In such a case make sure the device temperature has adjusted to the ambient temperature
before trying a new start-up.

4. If device is to be connected to other devices (e.g. via serial interface) the circuitry has to be designed

most carefully. Internal connection in third party devices (e.g. connection GND and earth) may result in
not-permissible voltages impairing or destroying the device or another device connected.

Warning:

  If device is operated with a defective mains power supply (e.g. short circuit from mains voltage

to output voltage) this may result in hazardous voltages at the device (e.g. at sensor socket or interface).

5. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to

be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer period of time.
In case of doubt, please return device to manufacturer for repair or maintenance.

6.

Warning:

 Do not use these product as safety or emergency stop device, or in any other application

where failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage.

1.2  Operation And Maintenance Advice 

Battery Operation

If ´bAt´ is shown in the secondary display the battery has been used up and needs to be replaced. The
device will, however, operate correctly for a certain amount of time. If ´bAt´ is shown in the upper display
the voltage is too low to operate the device; the battery has been completely used up.

Please note:  The battery has to be taken out, when storing device above 50°C.

We recommend to take out battery if device is not used for a longer period of time! 
The real time clock has to be set again after reconnect to the battery.

 

Mains Operation With Power Supply
Warning: 

When using a power supply please note that operating voltage has to be 10.5 to 12 V DC.

Do not apply overvoltage!! Cheap 12V-power supplies often have excessive no-load voltage.
We, therefore, recommend using regulated voltage power supplies. Trouble-free operation is guaranteed
by our power supply GNG10/3000.
Prior to connecting the power supply to the mains make sure that the operating voltage stated at the
power supply is identical to the mains voltage.

ƒ

Connecting/Changing Sensors

Do not use insuitable sensors. Connecting other devices/sensors as specificated may cause a damage to
the instrument and device/sensor! Switch off device before changing the sensor.
Connect sensors before switching on the device, otherwise the sensor may not be detected correctly.
When connecting the sensor the connector may not lock correctly. In such case take the plug not at the
casing but at the buckling protection at the end of the plug. If plug is entered correctly, it will slide in
smoothly. To disconnect sensors do not pull at the cable but at the plug (to open locking mechanism).

SIKA Dr.Siebert & Kühn GmbH & Co. KG ·• Struthweg 7–9 • 34260 Kaufungen • Germany • 

 +49 5605 803-0 • 

 +49 5605 803-54 • 

 [email protected] • www.SIKA.net 

Содержание MH3111

Страница 1: ...r s Manual Handheld Pressure Meter MH3111 as of Version 6 4 for MSD or MXD Pressure Sensors SIKA Dr Siebert K hn GmbH Co KG Struthweg 7 9 34260 Kaufungen Germany 49 5605 803 0 49 5605 803 54 info sika...

Страница 2: ...URE SENSORS 5 2 3 P OFF AUTO POWER OFF TIME 5 2 4 ADR BASE ADDRESS OF INTERFACE 5 2 5 OFFS ADJUSTUNG SENSOR ZERO DISPLACEMENT 5 2 6 SCAL ADJUSTUNG SENSOR SCALE 5 3 MEASURING OF WATER LEVEL DISPLAY UNI...

Страница 3: ...e product as safety or emergency stop device or in any other application where failure of the product could result in personal injury or material damage Failure to comply with these instructions could...

Страница 4: ...max key 2 shows the maximum of the measured value Pressing it again hides them To clear the max memory press key max for 2 seconds Min Memory Pressing min key 5 shows the minimum of the measured valu...

Страница 5: ...s the absolute pressure at sea level zero 2 3 P oFF Auto Power Off Time The device will be automatically switched off if no key is pressed no interface communication takes place for the time of the po...

Страница 6: ...ASYCONTROL Universal multi channel software EASYBUS RS485 or MH3000 operation possible for real time recording and presentation of measuring data of one MH3xxx device in the ACCESS data base format In...

Страница 7: ...System Messages Display Meaning What to do Low battery power device will only continue operation for a short period of time Replace battery Battery empty Replace battery Mains operation without batter...

Страница 8: ...for measuring values and for min max memories hold function etc as well as additional functional arrows Pushbuttons 6 membrane keys Interface Serial interface 3 5mm jack can be connected to RS232 or...

Отзывы: