Instruction Manual StackGuard
18
10266E/1
Figure
19: Terminal strip
for 85
.
.264 VAC
For information about control signals, please refer to Section 2 in the
Reference Handbook.
Terminals Use
Remarks
1.
20 .. 23
Connection to the
photometer
Wire colors as in connection plate.
4 - 5 - 6
Relay output 1
Relay output 1 can be freely config-
ured (
→
7 - 8 - 9
Relay output 2
The purge air heater is connected
via relay output 2 (
→
18 - 19
Output signal 1
0/4..20 mA, max. load 600
Ω
If unused, short-circuit these termi-
nals with a jumper!
2.
32 - 33
Output signal 2
0/4..20 mA, max. load 600
Ω
If unused, short-circuit these termi-
nals with a jumper!
10 .. 17
Inputs/outputs for
measuring range 1
→
Reference Handbook
26 .. 31
Inputs/outputs for
measuring range 2
→
Reference Handbook
16a
Remote triggering of
a sensor check
→
Reference Handbook
3.
24 - 25
Connection to bus
coupler
→
Reference Handbook
1 - 2 - 3
Mains voltage
85 .. 264 V; 47 .. 63 Hz or
24 VDC; 25 W
4.
34 - 35 -36 Purge air heater
→
Section 3.9
Terminal
assignment