background image

11

 2551 Magmeter 

• 

Conductivity is less than 20 

μ

S/cm.

• 

Electronic component failure.

• 

Loop device not scaled same as 
Magmeter.

• 

Range Jumper not placed correctly.

• Defective 

Magmeter

• 

Magmeter installed too close to 
upstream obstruction.

• 

Electrodes coated with solids. 

• 

Electrodes exposed to air bubbles.

• 

Electrical noise interference.

• 

New sensor, metal surface not 
properly conditioned.

• 

Output is not 0 when 

fl

 ow is stopped.

• 

Frequency, Digital or Current

 

output is erratic.

• 

Output is 22.1 mA.

• 

Electrode not adequately conditioned.

• 

Electrical noise interference.

• 

Vibration or other movement in pipe

 

causes magmeter to detect 

fl

 ow.

• Defective 

Magmeter

Symptom

Possible Cause

Possible Solution

• 

Frequency output is inoperative

• Digital 

(S

3

L) output is inoperative.

• 

Loop output is inoperative.

• 

Move the Magmeter upstream at least 
10 pipe diameters from obstruction.

• 

Clean the electrodes with soft cloth. 
Do not use abrasives.

• 

Eliminate air bubbles in the pipe.

• 

Remove the Magmeter and reinstall 
with the 

fl

 ow direction arrow on the 

sensor body pointed DOWNSTREAM.

• 

Modify grounding as required 
to protect the Magmeter from 
interference.

• 

Soak sensor overnight in 

fl

 uid.

11.2  

Troubleshooting

11. Maintenance

The 2551 Magmeter requires very little maintenance.  There are no user-serviceable components in the Magmeter.
• If 

the 

fl

 uid contains deposits and solids that may coat the electrodes, a regular cleaning schedule is recommended.

• 

Do not use abrasive materials on the metal electrodes.  Clean with soft cloth and mild detergent only.

• 

Use a cotton swab and mild detergent to remove deposits on the metal electrodes at the tip of the sensor.

11.1.  

Environmental  Recommendations:

• 

When used properly, this product presents no inherent danger to the environment.

• 

Please follow local ordinances when disposing of this or any product with electronic components.

• 

Soak sensor overnight in 

fl

 uid.

• 

Modify grounding to protect the 
Magmeter from interference.

• Set 

low 

fl

 ow cutoff higher.

• 

Return to factory for service.

• 

Use 3-0250 Setup tool to respan the 
Magmeter to match Loop device.

• 

Respan Loop device to match 
Magmeter.

• 

Set Range Jumper correctly.

• 

Return to factory for service.

• 

2551 is wrong model.

• 

Blue jumper not in correct position.

• 

Wiring is not correct.

• 

Frequency input to other 
manufacturer's 

fl

 ow instrument does 

not have pull-up resistor.

• 

Frequency model is 3-2551-11.

• 

Place blue jumper correctly.

 

(Sec. 5 pg. 4)

• 

Check wiring, make corrections.

• Install 

10k

Ω

 resistor. (Sec 8.1, pg. 6)

• 

Unsuitable application for Magmeter.

   

Return to factory for service.

• 

4-20 mA output is incorrect.

11.3 Troubleshooting with the RED and BLUE lights

No Lights:  

 

The power is off or the sensor is not connected

Solid Blue:  

 

The power is on but there is no 

fl

 ow in the pipe.

Blinking Blue:  

 

Normal operation, blink rate is proportional to the 

fl

 ow rate.

Alternating Red-Blue:  

Empty pipe indication (electrodes are not wet.)

Blinking Red:  

 

System errors (Electrical noise interference)

Solid Red:  

 

Instrument error (defective electronics component) 

1

2

3

4

D7

D6

Flow

Flow

If the 2551 detects an Empty Pipe:

• 

Frequency output will be locked to 0 Hz if electrodes are not wet.

• Digital 

(S

3

L) output will be locked to 0 if electrodes are not wet.

• 

4-20 mA will be locked to 4 mA if electrodes are not wet.

• 

Blue and Red LED indicators on the magmeter circuit will blink alternately if electrodes are not wet.

If the 2551 detects REVERSE FLOW:

• 

Frequency out cannot distinguish reverse 

fl

 ow from forward 

fl

 ow. The output will be the absolute value.

• Digital 

(S

3

L) output: Reverse 

fl

 ow results in 0 

fl

 ow rate displayed on 8900

• 

4-20 mA output can be spanned into negative 

fl

 ow range using the 3-0250 USB Setup Tool and software. (See section 9)

 

(example: 4-20 mA = -100 to +100 GPM)

Содержание 2551

Страница 1: ...two cable ports Use appropriate hardware to secure the 2551 from moisture intrusion One Liquid Tight Connector is included Sec 5 Pg 4 Topic Page 1 Quick Start Guide 1 2 Specifications 2 3 Installation Pipe Fittings 3 4 Selecting a Location 3 5 Hardware Configuration 4 6 General Installation and Grounding Tips 5 7 Wiring the Magmeter with 4 20 Loop 5 Topic Page 8 Wiring the Magmeter with Frequency ...

Страница 2: ... PBT Display window Polyamide Electrical Power Requirements 4 to 20 mA 21 6 to 26 4 VDC 22 1 mA max Frequency 5 to 26 4 VDC 15 mA max Digital S3 L 5 to 6 5 VDC 15 mA max Reverse polarity and short circuit protected Current output 4 to 20 mA Loop Accuracy 32 μA max error 25 C 24 VDC Isolation Low voltage 48 VAC DC from electrodes and auxilary power Maximum cable 300 m 1000 ft Error condition 22 1 m...

Страница 3: ...tance of straight pipe immediately upstream of the sensor Locating the sensor in a trap or where the flow is upward helps to protect the sensor from exposure to air bubbles when the system is in operation These configurations are not recommended because it is difficult to keep the pipe full Type Description 2 to 4 inch cut 1 7 16 inch hole in pipe Over 4 inch cut 2 1 8 inch hole in pipe 0 5 to 4 i...

Страница 4: ...too deeply and damage the Magmeter wiring 1 2 3 4 JP2 White Yellow Red Black Brown Blue Not used The factory connects the sensor cable to the termin inside the yellow cover This blue jumper selects frequency output or S L serial data output in the 3 2551 XX 11 Not used on 4 20 mA models 3 The user must connect output cables to this 4 terminal block JP2 is for factory use only MAKE NO CONNECTIONS U...

Страница 5: ...mps on metal pipes Fluid grounds must be in direct contact with the fluid and as near to the Magmeter as possible e The shield from the output cable must be terminated at the remote instrument ONLY This shield must not be connected at both ends f Connect an additional wire minimum AWG 14 2 08 mm2 from the remote instrument ground to the Magmeter ground terminal Grounding rings on plastic pipe Inst...

Страница 6: ... Jumper OFF S3 L OUT S L 3 3 4 8 Wiring the 3 2551 11 with Frequency or Digital S3 L output 1 2 3 4 5 24 VDC Ground Not used Frequency Out 2551 Magmeter 10KΩ Install a pull up resistor when connecting the 2551 Magmeter to other manufacturer s flowmeters 2551 Frequency Out to other Manufacturer s equipment Blue Jumper ON FREQ OUT 3 4 Frequency 1 2 3 4 5 24 VDC Ground Not used 9 8 7 Sensr Gnd SHIELD...

Страница 7: ...7 1 CPV8T015 325 56 86 01 92 15 348 78 SCH 80 PVC SADDLES FOR SCH 80 PVC PIPE 2 PV8S020 193 83 51 21 154 77 585 81 2 PV8S025 138 01 36 46 217 38 822 78 3 PV8S030 83 89 22 16 357 62 1353 60 4 PV8S040 40 88 10 80 733 88 2777 74 6 PV8S060 22 53 5 95 1331 85 5041 06 8 PV8S080 12 52 3 31 2395 41 9066 64 10 PV8S100 7 94 2 10 3778 75 14302 57 12 PV8S120 5 71 1 51 5256 69 19896 57 SCH 80 PVC SADDLES FOR S...

Страница 8: ...8 34 DN20 SFMT007 1158 05 305 96 25 91 98 05 DN25 SFMT010 749 09 197 91 40 05 151 58 DN32 SFMT012 439 51 116 12 68 26 258 36 DN40 SFMT015 248 93 65 77 120 52 456 16 DN50 SFMT020 146 85 38 80 204 30 773 26 PVC FITTINGS DIN ISO BS ANSI DN15 PVMT005 2067 76 546 30 14 51 54 91 DN20 PVMT007 1136 61 300 29 26 39 99 90 DN25 PVMT010 716 52 189 31 41 87 158 47 DN32 PVMT012 446 07 117 85 67 25 254 56 DN40 P...

Страница 9: ...5 4 CR4W040 36 84 9 73 814 34 3082 28 5 CR4W050 29 28 7 73 1024 70 3878 50 6 CR4W060 20 29 5 36 1478 26 5595 21 8 CR4W080 11 73 3 10 2557 72 9680 96 10 CR4W100 7 45 1 97 4028 83 15249 13 12 CR4W120 5 24 1 39 5722 73 21660 53 CARBON STEEL WELDOLETS ON SCH 40 PIPE 2 CS4W025 105 70 27 93 283 82 1074 27 3 CS4W030 70 68 18 67 424 45 1606 56 4 CS4W040 36 38 9 61 824 65 3121 30 5 CS4W050 29 28 7 73 1024 ...

Страница 10: ...K Factor K Factor 20 mA 20 mA IN TYPE Gallons Liters in GPM in LPM BRONZE TEES ON SCH 40 PIPE 1 BR4T010 582 34 153 86 51 52 194 99 1 BR4T012 330 54 87 33 90 76 343 53 1 BR4T015 254 76 67 31 117 76 445 71 2 BR4T020 157 36 41 58 190 64 721 58 COPPER TEES FITTING ON COPPER PIPE SCH K CUKT005 2459 19 649 72 12 20 46 17 CUKT007 1108 02 292 74 27 08 102 48 1 CUKT010 649 87 171 70 46 16 174 73 1 CUKT012 ...

Страница 11: ...follow local ordinances when disposing of this or any product with electronic components Soak sensor overnight in fluid Modify grounding to protect the Magmeter from interference Set low flow cutoff higher Return to factory for service Use 3 0250 Setup tool to respan the Magmeter to match Loop device Respan Loop device to match Magmeter Set Range Jumper correctly Return to factory for service 2551...

Страница 12: ... output 3 2551 P 0 1 2 Example Part Number Replacement parts and Accessories Mfr Part No Code Description 1220 0021 198 801 186 O ring FPM 1224 0021 198 820 006 O ring EPDM 1228 0021 198 820 007 O ring FFPM 3 2551 11 159 001 215 Magmeter electronics no display frequency or digital S3 L output 3 2551 12 159 001 216 Magmeter electronics no display 4 to 20 mA output 3 2551 P0 159 001 211 PP 316L SS D...

Отзывы: