Signet 2100 Turbine Sensor Operating Instructions

WARNING!

1.  Do not remove from pressurized lines.

2.   Never install sensor without O-rings.

3.  Confirm chemical compatibility before use.

4.  Do not exceed maximum temperature/pressure specifications.

5.  Wear safety goggles and faceshield during installation/service.

6.  Failure to follow safety instructions could result in severe personal injury.

Contents

1.  Installation/Mounting

2.  Sensor Wiring

3.  Calibration

4.  Specifications

5.  Parts and Accessories

1.  Install sensor with arrow pointing in the direction 

 

of flow.  All mounting angles acceptable.

2.  Slide union nut onto pipe or flexible tubing.

3.  Install end connectors.

4.  Hand tighten union nut to secure.

•  Suitable for clear or opaque fluids.  Suspended solids may 

cause mechanical failure.

•  Upstream/downstream mounting requirements: 

10x I.D. 

5x I.D. 

Inlet 

Outlet 

Flange 

15x I.D. 

5x I.D. 

Reducer 

40x I.D. 

5x I.D. 

2 x 90° Elbow 

 

50x I.D. 

Valve/Gate 

5x I.D. 

20x I.D. 

5x I.D. 

90° Elbow 

25x I.D. 

5x I.D. 

2 x 90° Elbow 

OK

OK

OK

Vertical flow is OK if the pipe 

remains full at all times.

64 mm

(2.5 in.)

33.3 mm

(1.3 in.)

43.2 mm

(1.5 in.)

12.2 mm

(.48 in.)

Pipe

End Connector

Union Nut

Reversible

electronics

O-rings 

(factory installed)

Press to release

Wiring Tips:

•  Use 2-conductor twisted-pair shielded cable (Belden 

#8451 or equivalent) for sensor cable splices up to 300 m 

(1000 ft) max.

•  Maintain cable shield through cable splice.

•  Route sensor cable away from AC power lines.

Input

Gnd.

10 kΩ 

+

5 to 24

VDC

-

Black  

Shield

Red

Other

instrument

Black (5 to 24 VDC)

Red (Signal out)

Shield (DC return)

Blk 

Power

Red 

Freq. input

Shld

Gnd

Signet

2100 Turbine

Flow Sensor

Other

Instrument

Signet

Instrument

Input

Gnd.

10 kΩ 

+

5 to 24

VDC

-

Black  

Shield

Red

Other

instrument

Black (5 to 24 VDC)

Red (Signal out)

Shield (DC return)

Blk 

Power

Red 

Freq. input

Shld

Gnd

Signet

2100 Turbine

Flow Sensor

Other

Instrument

Signet

Instrument

*3-2100.090-1*

3-2100.090-1  Rev. H 01/15   English

Sensor Model

Connection Option

K-Factors: Pulses/U.S. Gallon

Pulses/Liter

High Flow Sensors 3-2100-1H, 2H

1/2" Pipe

1725

456

1/2" Hose

1700

449

Low Flow Sensors 3-2100-1L, 2L

1/2" Pipe

12250

3236

1/4" Hose

12800

3382

3/8" Hose

13000

3435

1/2" Hose

12500

3303

Installation/Mounting

Sensor Wiring

Calibration