Signametrics
20
2.10.2 Frequency and Period Measurement
ACV Mode
•
Input Impedance
1 M
Ω
with < 300 pF
Frequency Range
2 Hz - 100 Hz
100 Hz-1 kHz
1 kHz-10 kHz
10 kHz-100 kHz 100 kHz-300 kHz
Resolution
1 mHz
10 mHz
100 mHz
1 Hz
1 Hz
Uncertainty is ±0.002% of
reading ± adder shown
4 mHz
20 mHz
200 mHz
2 Hz
5 Hz
Input Signal Range [1]
10% - 200%
of range
10% - 200%
of range
10% -200%
of range
10% - 200%
of range
45% -200%
of range
[
1] Input RMS voltage required for a valid reading. Do not exceed 250 V RMS input. For example, 10% -200%
of range indicates that in the 240 mVAC range, the input voltage should be 24 mV to 660 mV RMS.
ACI Mode
•
Input Impedance
10
Ω
in the 3 mA and 30 mA ranges, 0.1
Ω
in the 330 mA and 2.5 A ranges.
Frequency Range
2 Hz - 100 Hz
100 Hz-1 kHz
1 kHz-10 kHz
10 kHz-500 kHz
Resolution
1 mHz
10 mHz
100 mHz
1 Hz
Uncertainty
0.01% ±4 mHz
0.01% ±20 mHz
0.01% ±200 mHz
0.01% ±2 Hz
Input Signal Range,
2.4 mA, 240mA Ranges
[1]
10% -500%
of range
10% - 500%
of range
10% -500%
of range
10% - 500%
of range
Input Signal Range,
24 mA, 2.4 A ranges
50% -100%
of range
50% - 100%
of range
50% - 100%
of range
50% - 100%
of range
[1] Input current required to give a valid reading. For example, 10% -500% of range indicates that in the 3.3 mA
range, the input current should be 0.33 mA to 16.5 mA.
2.10.3 Duty Cycle Measurement
Frequency Range
2 Hz to 100 Hz
100 Hz to 1 kHz
1 kHz to 10 kHz
10 kHz to 100 kHz
Resolution
0.02%
0.2%
2%
20%
Typical Uncertainty is
±0.03% of reading ±
adder shown
0.03%
0.3%
3%
20%
Full scale reading
100.00 %
100.00 %
100.00 %
100.00 %
2.10.4 Pulse Width
± (% of r sec)
Polarity
Frequency range
Resolution
Width range
Typical
Uncertainty
Positive or negative pulse
widths
2 Hz to 100 kHz
1
µ
s
2
µ
s to 1 s
0.01 +/- 4
µ