![Signametrics SM2060 Скачать руководство пользователя страница 151](http://html1.mh-extra.com/html/signametrics/sm2060/sm2060_operators-manual_1268482151.webp)
151
Signametrics
The
SM60CAL.DAT
file is created by performing external calibration. The general calibration
algorithm consists of applying a zero value to the DMM followed by a value of 2/3
P
rd
P
of the top of each
range. Calibration of your SM2060/64 is best performed using calibration software available from
Signametrics.
When using multiple DMMs in a single chassis, the
SM60CAL.DAT
file must have a calibration record
for each DMM. You can combine the unique calibration records of each DMM into one
SM60CAL.DAT
file using any ASCII text editor such as “notepad.exe”.