background image

2022.05

Automatic

 

Intelligent

 

OPTICALFIBER

 

FUSION

 

SPLICER

AI-6C

Содержание AI-6C

Страница 1: ...2022 05 Automatic Intelligent OPTICALFIBERFUSIONSPLICER AI 6C ...

Страница 2: ...UCTION INTRODUCTION TO SPLICE INTERFACE CHAPTER Ⅱ SOFTWARE DOWNLOAD SCREEN MENU FUNCTION ON OFF CHAPTERⅢ ARC CALIBRATION CHAPTER Ⅳ FIBER STRIPPING AND PUT INTO HOLDER CHAPTER Ⅴ REPLACE ELECTRODE CHAPTER Ⅵ ACTIVATE ELECTRODE CHAPTER Ⅶ SMART LOCK CHAPTER Ⅷ DAILY MAINTENANCE 1 2 5 9 10 13 14 16 17 19 21 23 25 ...

Страница 3: ...amage To The Fusion Splicer AC DC Power Adapter Input AC Voltage Of 100 240v 50 60hz If Input Voltage Exceeds This Range May Cause Permanent Damage To The Adapter Internal Lithium Battery There Is A Lithium ion Battery Cells In The Fusion Splicer The Use Of Other Batteries May Damage The Fusion Splicer And Jeopardize Personal Safety For Safety Sake Lithium Battery Pack Can Not Be Disassembled To P...

Страница 4: ...tion three to activate the battery Notice 1 Optical fiber fusion splicer is used for welding quartz glass fiber please do not use this instrument for other purposes Please read this manual carefully before use 2 Pay attention to clean the V groove and lens in the process of use to clean V groove you can use an art knife to draw back and forth along the fiber groove and then brush with a brush Clea...

Страница 5: ...e phone again zhuoshizhangzhongbao firmware updates are more stable faster and can force a firmware refresh When normal upgrades fail it s easier to succeed with this feature 2 Self check of optical fiber fusion splicer in the shutdown state press and hold the continue button and the power button at the same time and the machine enters the self check mode Usage When the welding machine is not work...

Страница 6: ... in the state of power off and wait for 2 hours before starting up to check Battery cannot be normally started can only plug the adapter to start up after starting up long press the Magnifier button about 8 seconds after the buzzer will tick prompt and then manually shut down plug in the adapter in the state of power off wait 2 hours before starting the machine Usage The welding machine has abnorm...

Страница 7: ...atically target value in the Splicing settings menu set focus target value ARC calibration According to the altitude temperature humidity and other conditions of the user match the most suitable arc discharge value to minimize the splIcIng lOSS See chapter 4 discharge correction Smart lock purpose To Facilitate Management Or Owner s Effective Management Of The Splicer Machine s Use Period Or The N...

Страница 8: ...p of the electrode and remove debris Restore the factory Settings in the device s fusion setting Restart welding machine and zhuoshizhangzhongbao re discharge correction If the fault continues enter the welding machine self check mode to check whether item 11 is wrong The left and right optical fibers do not match the specifications or types of optical fibers on both sides are Intelligent prompt a...

Страница 9: ... FUNCTION ON OFF Turn on the phone bluetooth open the phone app set the machine panel menu after establishing the connection and machine will update at the same time The splicing operation does not require phones the phone App is only for machine menu settings and splicer machine to store data when the data is exported to phone App the users can also get support through the APP and after sales ser...

Страница 10: ...ch as single mode and multi mode switching The procedure for ARC Calibration is as follows Step 1 Please turn on the splicer machine then login Signalfire 2 on your phone connect the Bluetooth open the Arc Calibration See picture Step 2 After open the ARC Calibration the icon of Normal Mode will change to the icon for ARC fiber place the fiber to fiber holder and close the o Calibration Then pleas...

Страница 11: ...are fibe Cable bare fibe Note 1 After all the optical fibers are stripped and placed according to the above length the front end should be at least completely pressed by the fiber compression board on the right of the blade otherwise it will be continuously cut 2 Please keep the four pressure fiber rubber pads clean no waste fiber and other sundries avoid contact with alcohol and other corrosive l...

Страница 12: ...ly see the location in picture then remove up the electrode cover 2 The cap size of two electrodes is different wider head is for the electrode which is close to screen direction 3 Take away the cover as shown in picture 4 Remove the lamp cap on both sides with tweezers wider part of latch 5 Screw out the screws with a screwdriver 6 use a screwodriver to push out the latch and remove the electrode...

Страница 13: ...tware 1 Open the signalfire2 APP software click on login 2 Please log in directly if you have an account otherwise please registe first 3 Click Device unconnected after login is successful 4 Please select the corresponding Bluetooth number and click on if you can not find the bluetooth number please check whether the Bluetooth or the machine power is turned on 5 Connecting 6 Connected successfully...

Страница 14: ...r 2 After Binding Is Clicked this Mobile Phone Account Has Unique Control Over This Splicer AI 6C V02420145358411 There is no phone attached to the machine binding device connected A Device Can Only Be Bound To One Mobile Account At The Same Time To Bind Other Mobile Phones You Need To Unbind Current Account First 3 Click on the settings icon to enter 4 Save after personalization settings 5 Click ...

Страница 15: ...ton swab to wipe from the middle of the lens do a circular motion until the edges of the lens spin out Then wipe with a clean dry cotton swab to remove extra alcohol 3 Turn on the power make sure that the display is not visible dust and stripes Ⅱ Prevent Strong Shock or Vibration When you need to move or transport the fusion splicer you should handle with care and gently In addition do not forget ...

Страница 16: ...i mode fiber splice 1 Fiber end with burr not flat 2 Fiber quality is not good 1 Increase the cleaning voltage 2 Re cleave fiber 3 Replace to qualified fiber or cut off a part of the existing fiber and try again DAILY MAINTENANCE Troubleshooting Reason Solutions Splicing points have lateral shadows 1 Fiber core does not match type or core diameter is different 2 Multimode fiber appear very light s...

Отзывы: