Laboratory Centrifuge Sigma 6-16S
5 Set-up and connection
28 / 91
Version 07/2012, Rev. 1.22 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0702002
5.2.2
Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp L or B fuses that
are to be provided by the customer.
NOTE
To ensure safe disconnection in the event of a fault, an AC/DC-sensitive
RCD (residual current device) must be integrated in the wiring system of
the building.