Laboratory Centrifuge Sigma 6-16S
4 Storage and transport
26 / 91
Version 07/2012, Rev. 1.22 of 08/05/2020
• sb
Translation of the original operating manual, part no. 0702002
4.4
Packaging
The centrifuge is packaged in a wooden crate.
• After taking off the lid, remove the side panels.
• Remove the packaging material.
• Lift the centrifuge upwards with a lifting device to lift it safely. When
lifting the centrifuge, always reach under the centrifuge from the side.
CAUTION
The centrifuge weighs approx. 116 kg!
• Retain the packaging for any possible future transport of the centrifuge.
4.5
Transport safety device
The transport safety device consists of two plastic screws which are located
at the bottom panel.
CAUTION
The transport safety device must be removed prior to start-up because the
screws lock the motor bearings!
Removal
• Lift the centrifuge upwards at the front side. Always reach under from
the side.
• Put a suitable object, e.g. a wooden block, between the table and centri-
fuge. The two plastic screws can now be seen at the bottom panel.
1
Locking screws
Fig. 6: Locking screws (transport safety device)
• Unscrew the screws by hand anti-clockwise.
• Retain the transport safety device for the possibility of the return of the
centrifuge.